Six weeks after a measurement, someone asks why your amplifier's gain was 2 % low. If the answer is in your notebook (the meter you used, its calibration date, the room temperature, the supply setting, the raw readings and the fact that you replaced R4 halfway through), you can answer in a minute. If it is not, the measurement is lost, however carefully it was made. This lesson covers the engineering notebook, a Git repository for lab work, and the lab report, whose first version is this course's milestone from the owner's roadmap (Stage 0): objectives, circuit diagram, equipment list, measurements, uncertainty, results and conclusions.

The notebook: a record, not a diary

p.142026-11-28 Cell voltage vs meter spec Objective: an AA cell vs the meter's accuracy. Meter: 6000-count DMM, s/n 21-0457, cal. due 2027-03, DCV +/-(0.5 % + 3 counts) Conditions: 22.5 C, 48 % RH; cell unused. #rangereading 16 V1.587 V 26 V1.578 V 1.587 V (digits swapped; TA) 360 V1.59 V Result: 1.587 +/- 0.011 V on the 6 V range; +/- 0.038 V on the 60 V range. Readings 10:05-10:07, cell at room temp 1 h. Signed: TAWitness: -

Bound paper notebook. Numbered pages, never torn out; ink; every entry dated; blank space struck through so nothing can be added later. Mistakes are crossed out with a single line, so they stay readable, with the correction, the reason and your initials next to them. In industry and research a witnessed, signed notebook can be evidence of when an idea or result existed.

Electronic lab notebook (ELN). A notebook application or a set of plain-text files in a Git repository. It keeps timestamps and an audit trail of every change, holds raw data files and scripts directly, and is searchable. The same rule applies: you add and correct, you never silently overwrite.

What to record for every session:

  • date, time, page or entry number; your name;
  • the objective in one sentence;
  • the setup: schematic or a photo of the wiring, component values as measured where it matters;
  • the equipment: make and model class, serial or asset number, the range used, calibration due date;
  • conditions: room temperature, humidity if relevant, warm-up time, supply settings including the current limit;
  • raw data exactly as read, with units, before any calculation; the time of each reading if it might drift;
  • what you changed and why, what surprised you, and anything that went wrong.

A Git repository for lab work

Treat each project as a repository. Raw data goes in once and is never edited; processing scripts read it and write results.

text
divider-lab/
  README.md            objective, how to reproduce, who did what
  notebook/            dated entries: 2026-11-28-cell.md, 2026-11-30-divider.md
  data/raw/            CSV files straight from the bench (never edited)
  data/processed/      written by scripts, can be regenerated
  sim/                 SPICE decks (divider.cir)
  scripts/             analysis.py, plots.py
  report/              report.md and its figures
  .gitignore

A minimal .gitignore keeps generated and private files out:

text
__pycache__/
*.pyc
.ipynb_checkpoints/
data/processed/
*.raw
*.log
.env

The commands (from the Git course) are the everyday ones:

bash
git init divider-lab
cd divider-lab
git add README.md data/raw/2026-11-28-cell.csv notebook/2026-11-28-cell.md
git commit -m "Cell voltage on two ranges, raw data and notes"

Commit small and often, with messages that say what was measured, not "update".

The lab report

Section What it contains
Title, author, date and the notebook pages or commit the work comes from
Objectives what question the experiment answers
Theory and prediction the hand calculation and simulation, with their uncertainty
Circuit diagram an IEC schematic with every value and the measurement points marked (Reading Schematics)
Equipment list instruments with model class, serial or asset number, ranges and calibration dates; components with tolerances
Method the steps, enough for someone else to repeat them; the safety setup
Measurements raw data tables with units, and conditions
Uncertainty the budget: Type A and Type B inputs, combined and expanded (Measurement Uncertainty)
Results prediction versus measurement, with a stated test of agreement
Conclusions the answer to the objective, what limited the result, what to do next

A good test of agreement between a prediction xp±Upx_p \pm U_p and a measurement xm±Umx_m \pm U_m (both expanded with k=2k = 2) is the normalised error

En=∣xm−xp∣Up2+Um2E_n = \frac{\lvert x_m - x_p \rvert}{\sqrt{U_p^2 + U_m^2}}

used in laboratory proficiency testing (defined in ISO 13528:2022, the statistical companion to ISO/IEC 17043:2023). En≤1E_n \le 1 means the two agree within their uncertainties; En>1E_n > 1 means something is not understood.

Worked example: does the divider agree?

Given: prediction Vout=3.837±0.030 VV_\text{out} = 3.837 \pm 0.030\ \text{V} (resistor tolerances), measurement 3.821±0.022 V3.821 \pm 0.022\ \text{V} (meter specification).

Find: EnE_n and the verdict.

Solve:

En=∣3.821 V−3.837 V∣(0.030 V)2+(0.022 V)2=0.016 V0.0372 V=0.43E_n = \frac{\lvert 3.821\ \text{V} - 3.837\ \text{V} \rvert}{\sqrt{(0.030\ \text{V})^2 + (0.022\ \text{V})^2}} = \frac{0.016\ \text{V}}{0.0372\ \text{V}} = 0.43

Check: EnE_n is dimensionless, as it must be, and below 1: the measurement agrees with the prediction. The difference of −0.42 %-0.42\,\% is not a result to explain away; it is inside what the components and meter allow.

A script can write the results table for the report straight from the data, so the numbers in the report are never retyped:

python · try it yourself
import math

# Predicted (hand calculation or SPICE) and measured values, each with its expanded uncertainty U (k = 2).
# Values are kept as text so the table shows exactly the digits written in the notebook.
rows = [
    # quantity,       unit, predicted, U_pred,  measured, U_meas
    ("V(out)",        "V",  "3.837",   0.030,   "3.821",  0.022),
    ("I",             "mA", "0.816",   0.008,   "0.812",  0.009),
    ("V(cell)",       "V",  "1.60",    0.05,    "1.587",  0.011),
    ("V(out), 1M/1M", "V",  "5.00",    0.05,    "4.762",  0.027),
]
print("| Quantity | Predicted | Measured | Difference | E_n | Verdict |")
print("|---|---|---|---|---|---|")
for name, unit, p, Up, m, Um in rows:
    pv, mv = float(p), float(m)
    En = abs(mv - pv) / math.sqrt(Up**2 + Um**2)      # normalised error
    verdict = "agree" if En <= 1 else "investigate"
    print(f"| {name} | {p} +/- {Up} {unit} | {m} +/- {Um} {unit} | {100 * (mv - pv) / pv:+.2f} % | {En:.2f} | {verdict} |")
| Quantity | Predicted | Measured | Difference | E_n | Verdict |
|---|---|---|---|---|---|
| V(out) | 3.837 +/- 0.03 V | 3.821 +/- 0.022 V | -0.42 % | 0.43 | agree |
| I | 0.816 +/- 0.008 mA | 0.812 +/- 0.009 mA | -0.49 % | 0.33 | agree |
| V(cell) | 1.60 +/- 0.05 V | 1.587 +/- 0.011 V | -0.81 % | 0.25 | agree |
| V(out), 1M/1M | 5.00 +/- 0.05 V | 4.762 +/- 0.027 V | -4.76 % | 4.19 | investigate |

The last row is the high-resistance divider from The Digital Multimeter: En=4.19E_n = 4.19, so the prediction is missing something, here the meter's 10 MΩ10\ \text{M}\Omega loading. Writing "investigate" and then explaining it is exactly what the report is for.

Guided lab: your first report

Hand calculation. For the cell measured in the notebook figure, the example meter's limit is 0.005×1.587+0.003=0.011 V0.005 \times 1.587 + 0.003 = 0.011\ \text{V} on the 6 V6\ \text{V} range and 0.038 V0.038\ \text{V} on the 60 V60\ \text{V} range.

Simulation. Put your readings into the script above (or a CSV file it reads) and generate the results table.

What you would measure. One fresh AA cell, three readings on the 6 V6\ \text{V} range and one on the 60 V60\ \text{V} range, plus the room temperature: expect agreement between readings within a count or two (about 1.551.55 to 1.65 V1.65\ \text{V} for a fresh alkaline cell), and both ranges agreeing within their limits. Then write a two-page report with every section of the table above, commit it with the raw data, and ask someone to repeat the measurement from your report alone. Whatever they could not reproduce is what the report is missing.

Hazards and safe setup.

  • A single 1.5 V1.5\ \text{V} cell and a multimeter on a voltage range: the only hazard is a short circuit if the leads are in the A jack.
  • Do not open, heat or charge cells to "get a better reading".

Engineering ethics in records

Professional codes (for example the IEEE Code of Ethics and national engineering registration rules) ask engineers to be honest and realistic in stating claims based on available data. In a notebook and report that means:

  • record what happened, including failed runs and outliers; never drop a reading without recording that and why;
  • never change raw data; corrections are new, dated entries;
  • state the uncertainty and the conditions, so nobody over-trusts the number;
  • credit other people's work, data and code, and say which parts were simulated, assumed or taken from a datasheet.

Pitfalls

  • Writing results on loose paper "to copy neatly later".
  • Recording only the processed result, not the raw readings.
  • No equipment serial numbers or calibration dates: the result cannot be traced.
  • Editing a CSV file of raw data by hand.
  • A conclusion that restates the method instead of answering the objective.

Practice

  1. You notice a reading was written as 3.2813.281 instead of 3.821 V3.821\ \text{V}. How do you correct it in a bound notebook?
  2. Prediction 10.0±0.2 V10.0 \pm 0.2\ \text{V}, measurement 10.5±0.2 V10.5 \pm 0.2\ \text{V}. Compute EnE_n.
  3. Which folder of the repository layout should never be edited by hand?
Answer
  1. One line through 3.2813.281, the correct value 3.821 V3.821\ \text{V} beside it with the reason (digits transposed), the date and your initials.
  2. 0.5/0.04+0.04=0.5/0.283=1.770.5/\sqrt{0.04 + 0.04} = 0.5/0.283 = 1.77: they disagree; investigate.
  3. data/raw/.

Test yourself

How do you correct a mistake in a bound lab notebook?

  1. Erase it neatly
  2. Tear out the page
  3. Strike it through with a single line and write the correction, reason and initials
  4. Cover it with correction fluid

Which item belongs in the equipment list of a lab report?

  1. The meter's colour
  2. The meter's model class, serial or asset number, range used and calibration date
  3. Only the meter's brand
  4. Nothing: equipment does not affect results

A prediction and a measurement give En=0.4E_n = 0.4. This means

  1. they disagree
  2. they agree within their combined expanded uncertainties
  3. the measurement is 40 % off
  4. the uncertainty is too large to say

Why keep raw data in a separate folder that scripts only read?

  1. Git cannot track processed files
  2. So every result can be regenerated from untouched originals
  3. To save disk space
  4. Raw data is private