The Digital Multimeter
You measure a resistor divider fed from . Two equal resistors, so the middle node must sit at . Your meter says . Is the meter broken, are the resistors wrong, or did the act of measuring change the circuit? A digital multimeter (DMM) is the first instrument on every bench, and it is also the one people trust too much. This lesson covers what it measures, how to read its accuracy specification, and the two ways it disturbs the circuit: loading on the voltage ranges and burden voltage on the current ranges.
What a multimeter does
| Function | Connection | What the meter puts into the circuit |
|---|---|---|
| DC and AC voltage | in parallel with the part, leads in COM and V | a large input resistance, typically |
| DC and AC current | in series, the circuit is opened and the meter closes it; leads in COM and mA or A | a small shunt resistance, so a small voltage drop (the burden voltage) |
| Resistance, continuity, diode test | across an unpowered part | a small test current from the meter's own battery |
| Capacitance, frequency, temperature (on many meters) | across the part, or a thermocouple | depends on the function |
Inside, every function ends as a DC voltage presented to an analogue-to-digital converter. Current is turned into voltage by a shunt resistor, resistance by a known test current. The display has a fixed number of counts: a 6000-count meter shows at most 5999 (or 6000) on any range, so on its range the last digit is , and on the range it is . That last digit is the resolution. It is not the accuracy.
Note: AC voltage and current readings depend on whether the meter is true RMS or average-responding (calibrated for sine waves only). The difference is explained in Alternating Current and RMS Values.
The same connections are in this lab with ideal meters (it writes voltage as , as School Physics does). Change and watch the ammeter follow :
At and it reads . Real meters are not ideal, and the rest of the lesson is about how far from ideal they are.
Reading an accuracy specification
Handheld meters state DC accuracy as
where is a percentage of the reading and is a number of counts (least significant digits). The spec holds for a stated temperature range (often ), humidity and time since calibration (often one year). Outside that, the manufacturer adds a temperature coefficient. The counts term matters most for small readings, which is why you choose the lowest range that does not overflow.
Worked example: a cell on two ranges
Given: an example 6000-count meter with DC-volts accuracy . An AA alkaline cell reads on the range (resolution ).
Find: the interval that contains the true voltage, and the same for a reading of on the range.
Solve:
So the cell is (between and ) on the right range, but only on the range.
Check: the percentage term is the same on both ranges (); only the counts term grew tenfold. That is the reason for the "lowest range" rule. These limits are a manufacturer's guarantee, so treat them as the half-width of a rectangular distribution when you build an uncertainty budget (Accuracy, Precision and Measurement Uncertainty).
The same calculation for several readings:
# Example spec of a 6000-count meter, DC volts: +/-(0.5 % of reading + 3 counts)
resolution = {"6 V": 0.001, "60 V": 0.01}
def dcv_limit(reading, rng, pct=0.5, counts=3):
return pct / 100 * reading + counts * resolution[rng]
for reading, rng in [(1.587, "6 V"), (0.150, "6 V"), (12.06, "60 V"), (1.59, "60 V")]:
u = dcv_limit(reading, rng)
print(f"{reading:7.3f} V on the {rng:>4} range: +/- {u:.4f} V = {100 * u / reading:.2f} % of reading") 1.587 V on the 6 V range: +/- 0.0109 V = 0.69 % of reading
0.150 V on the 6 V range: +/- 0.0037 V = 2.50 % of reading
12.060 V on the 60 V range: +/- 0.0903 V = 0.75 % of reading
1.590 V on the 60 V range: +/- 0.0379 V = 2.39 % of reading
Loading: the voltmeter is a resistor
A voltmeter on a DC range behaves like an ideal voltmeter in parallel with its input resistance , typically (check your meter's manual; some ranges and some meters differ). Seen from the meter, any circuit is a source behind a resistance (its Thévenin equivalent), so the meter reads
For a divider of two equal resistors , . With the error is , far below the meter's own accuracy. With , and the error is : the from the opening question. Nothing was broken.
def parallel(a, b):
return a * b / (a + b)
Vs, R_in = 10.0, 10e6 # source volts, meter input resistance in ohms
for R in [1e3, 10e3, 100e3, 1e6]:
lower = parallel(R, R_in) # the bottom resistor with the meter across it
v = Vs * lower / (R + lower)
print(f"R = {R / 1e3:6.0f} kohm: meter reads {v:.4f} V, error {100 * (v - 5) / 5:+.3f} %")R = 1 kohm: meter reads 4.9998 V, error -0.005 %
R = 10 kohm: meter reads 4.9975 V, error -0.050 %
R = 100 kohm: meter reads 4.9751 V, error -0.498 %
R = 1000 kohm: meter reads 4.7619 V, error -4.762 %
A real SPICE simulator agrees. Node a is the divider alone, node b the same divider with a resistor standing in for the meter (the deck format is explained in The EE Software Toolchain):
cat > load.cir <<'EOF'
* 1M/1M divider from 10 V: alone (node a) and read by a 10M meter (node b)
V1 in 0 DC 10
R1 in a 1meg
R2 a 0 1meg
R3 in b 1meg
R4 b 0 1meg
Rm b 0 10meg
.control
op
print v(a) v(b)
.endc
.end
EOF
ngspice -b load.cir 2>&1 | grep -E '^v\('v(a) = 5.000000e+00
v(b) = 4.761905e+00
Burden voltage: the ammeter is a resistor too
On a current range the meter inserts a shunt resistance , and the current through it drops the burden voltage . In a supply feeding , the ideal current is . With an example shunt of on the mA range the current becomes () and the meter drops . In low-voltage circuits (a logic rail, a sensor bias) this is a real error; manuals give the burden per range, often in mV per mA.
The current inputs are fused (on good meters with high-breaking-capacity fuses) because the most common meter accident is putting the probes across a voltage source while the leads are still in the A or mA jack: that is a near short circuit through the shunt.
Guided lab: a cell and a high-resistance divider
Hand calculation. Fresh AA alkaline cells read about to at no load (typical; check the cell's datasheet). A / divider from gives unloaded and with a meter.
Simulation. The Python loop and the SPICE session above.
What you would measure.
- The cell on the range, for example , quoted as with the example spec. Record the room temperature and the meter's model and calibration date in your notebook (The Lab Notebook).
- The divider, fed from a bench supply set to with a current limit. Expect , within about from the meter and about from resistors: clearly below . Repeat with resistors: now expect within the same limits.
- Measure yourself: put a known resistor in series with the meter across the supply. A meter reads about .
Hazards and safe setup.
- Low voltage only: a battery or a current-limited bench supply at DC. Never connect this lesson's setups to mains.
- Move the red lead back to the V jack before measuring voltage; a lead left in A across a supply blows the fuse or worse.
- Measure resistance and continuity only on unpowered, discharged circuits.
- A meter's CAT rating (IEC 61010-1, 3rd edition 2010 + A1:2016) matters for mains-connected work, which this course does not do; see Safe Bench Practice for what the categories mean.
Pitfalls
- Reading resolution as accuracy: four digits do not mean four correct digits.
- Measuring a high-impedance node (megohms, a CMOS input, a filter capacitor through a large resistor) and blaming the circuit for the meter's loading.
- Leaving the meter on a current range: the next voltage measurement is a short circuit.
- Measuring resistance in-circuit: other paths in parallel lower the reading.
- Using an average-responding meter on a non-sinusoidal waveform.
Practice
- A reading of on the range of the example meter: what is the limit ?
- A node with is read with a meter. What is the loading error in percent?
- A rail feeds ; the mA range has a shunt. What current does the meter read?
Answer
- , so .
- .
- instead of ; the burden voltage is .
Test yourself
A 6000-count meter on its 6 V range shows a resolution of
Why should you pick the lowest range that does not overflow?
- It makes the meter's input resistance larger
- The counts part of the accuracy spec is smaller on a lower range
- Higher ranges are not calibrated
- It reduces the burden voltage
A meter reads a node whose Thevenin resistance is . The reading is
- about 4.8 % low
- about 4.8 % high
- about 0.5 % low
- exact
What is the burden voltage of an ammeter?
- The largest voltage it can measure
- The voltage across its shunt resistance while current flows
- The voltage of its internal battery
- The voltage at which its fuse blows