Accuracy, Precision and Measurement Uncertainty
A measurement without an uncertainty is only half a result. "The resistor dissipates " invites the question: how sure are you of the last three digits? Engineers state results as a value with an interval, , and say how the interval was obtained. This lesson defines the words (accuracy, precision, repeatability, resolution), then applies the international method for evaluating uncertainty, the Guide to the Expression of Uncertainty in Measurement (GUM, published as JCGM 100:2008), to a voltage reading and a power calculated from it.
The School Physics course Lab Skills covers the school methods (half the range, percentage uncertainties, error bars); this lesson is the engineering version.
Four words that are not synonyms
| Word | Meaning (after the International Vocabulary of Metrology, JCGM 200:2012) | Improved by |
|---|---|---|
| accuracy | closeness of a result to the true value; needs both small bias and small scatter | calibration, a better instrument |
| trueness (low bias) | closeness of the average of many results to the true value | calibration, correcting systematic effects such as loading |
| precision | closeness of repeated results to each other | averaging, better technique, shielding |
| repeatability | precision under the same conditions, same operator and setup, short time | |
| reproducibility | precision across different operators, instruments, labs or days | |
| resolution | the smallest change the instrument can show (one count) | a higher-resolution instrument; it does not make the reading more accurate |
The GUM method in five steps
- Model the result as a function of the inputs, for example .
- Find a standard uncertainty (one standard deviation) for every input.
- Type A (from statistics of repeated readings): readings with sample standard deviation give for their mean.
- Type B (from any other information: specifications, calibration certificates, resolution). If all you know is that the value lies within with no value more likely than another, use a rectangular distribution: . A display resolution means .
- Combine: with sensitivity coefficients (inputs assumed independent). For a product of powers it is simpler in relative terms:
- Expand: with coverage factor , which gives an interval of roughly coverage when the combined distribution is close to normal.
- Report: value, , , rounded so has one or two significant figures and the value ends at the same decimal place.
Worked example: power in a resistor
Given: ten readings of the voltage across a , resistor on the range of an example meter with accuracy and resolution : mean , sample standard deviation .
Find: and its expanded uncertainty ().
Solve:
Result: ().
Check: the meter specification dominates the voltage budget; averaging more readings would shrink but barely change the answer. The voltage enters squared, so its relative uncertainty counts twice. A Monte Carlo simulation (sampling every input from its distribution, GUM Supplement 1, JCGM 101:2008) should give the same :
import numpy as np
readings = np.array([4.983, 4.991, 4.985, 4.990, 4.982, 4.988, 4.992, 4.984, 4.989, 4.986]) # volts
n = len(readings)
V = float(readings.mean())
s = float(readings.std(ddof=1))
u_A = s / np.sqrt(n) # Type A: standard uncertainty of the mean
a_spec = 0.005 * V + 3 * 0.001 # Type B: example meter spec, half-width in volts
u_spec = a_spec / np.sqrt(3) # rectangular distribution
u_res = (0.001 / 2) / np.sqrt(3) # display resolution
u_V = np.sqrt(u_A**2 + u_spec**2 + u_res**2)
R, u_R = 100.0, 1.0 / np.sqrt(3) # 100 ohm +/- 1 %, rectangular
P = V**2 / R
u_P = P * np.sqrt((2 * u_V / V)**2 + (u_R / R)**2)
print(f"V = {V:.4f} V, s = {s:.4f} V, u_A = {u_A:.4f} V, u_spec = {u_spec:.4f} V, u_res = {u_res:.4f} V")
print(f"u_c(V) = {u_V:.4f} V")
print(f"P = {P:.4f} W, u_c(P) = {u_P:.4f} W, U(k=2) = {2 * u_P:.4f} W")
rng = np.random.default_rng(1)
N = 200_000
V_mc = V + rng.normal(0, u_A, N) + rng.uniform(-a_spec, a_spec, N) + rng.uniform(-0.0005, 0.0005, N)
R_mc = rng.uniform(99.0, 101.0, N)
P_mc = V_mc**2 / R_mc
print(f"Monte Carlo: P = {P_mc.mean():.4f} W, u(P) = {P_mc.std():.4f} W")V = 4.9870 V, s = 0.0035 V, u_A = 0.0011 V, u_spec = 0.0161 V, u_res = 0.0003 V
u_c(V) = 0.0162 V
P = 0.2487 W, u_c(P) = 0.0022 W, U(k=2) = 0.0043 W
Monte Carlo: P = 0.2487 W, u(P) = 0.0022 W
Guided lab: repeated readings and a fitted resistance
Hand calculation. For the school "half the range" rule, ten readings that span give . The GUM Type A value for the same data is , about five times smaller for a normal spread of ten readings, because it describes the uncertainty of the mean, not of one reading.
Simulation 1. Ten readings of a rod on a vernier caliper, scattered by . With these settings the lab ends at by the half-range rule. Watch how the mean settles as readings arrive:
Simulation 2. A resistance from a fitted – line (the lab writes for voltage). Six points around a true with error bars: the least-squares gradient is , and the steepest and shallowest lines give . Try other data sets: the true value stays inside the interval most of the time, which is what an uncertainty promises.
What you would measure. Set a bench supply to about (current limit ) across a , , metal-film resistor and take ten voltage readings over a minute. Expect them to agree within a few millivolts (repeatability), a mean near , and a budget like the worked example: near , . If a second meter disagrees by more than both expanded uncertainties combined, one of them is outside its specification or a systematic effect (loading, a bad lead) is present.
Hazards and safe setup.
- DC from a current-limited supply; no mains.
- The resistor dissipates and gets warm; let it settle before reading (its value drifts with temperature, another input to your budget).
- Leads in the V jack, not A.
Pitfalls
- Quoting the resolution as the uncertainty.
- Using instead of for the uncertainty of a mean (or the reverse, when the question is about a single future reading).
- Adding uncertainties linearly instead of in quadrature (overestimates) or forgetting the sensitivity coefficient (the factor 2 for ).
- Treating a manufacturer's limit as a standard deviation ( for a rectangular limit).
- Too many digits: is not more accurate than .
Practice
- A spec says . What standard uncertainty do you assign, and why?
- with and . What is ?
- Twenty readings have . What is of their mean?
Answer
- : a limit with no other information is modelled as a rectangular distribution.
- (both exponents have magnitude 1).
- .
Test yourself
A meter shows six digits but is out of calibration by 2 %. Its readings are
- accurate and high resolution
- high resolution, not accurate
- precise, so accurate
- Type A uncertain
The standard uncertainty for a rectangular limit is
For , a relative uncertainty in contributes
What does the expanded uncertainty with give?
- The maximum possible error
- An interval of roughly 95 % coverage when the distribution is near normal
- Two standard deviations of a single reading only
- The instrument's resolution