Home › EE Foundations: Units, Measurement, Safety & the Lab Notebook › Accuracy, Precision and Measurement Uncertainty

Accuracy, Precision and Measurement Uncertainty

A measurement without an uncertainty is only half a result. "The resistor dissipates 0.2487 W0.2487\ \text{W}" invites the question: how sure are you of the last three digits? Engineers state results as a value with an interval, P=0.249±0.004 WP = 0.249 \pm 0.004\ \text{W}, and say how the interval was obtained. This lesson defines the words (accuracy, precision, repeatability, resolution), then applies the international method for evaluating uncertainty, the Guide to the Expression of Uncertainty in Measurement (GUM, published as JCGM 100:2008), to a voltage reading and a power calculated from it.

The School Physics course Lab Skills covers the school methods (half the range, percentage uncertainties, error bars); this lesson is the engineering version.

Four words that are not synonyms

accurateand precise precise,biased unbiased,not precise neither
Word Meaning (after the International Vocabulary of Metrology, JCGM 200:2012) Improved by
accuracy closeness of a result to the true value; needs both small bias and small scatter calibration, a better instrument
trueness (low bias) closeness of the average of many results to the true value calibration, correcting systematic effects such as loading
precision closeness of repeated results to each other averaging, better technique, shielding
repeatability precision under the same conditions, same operator and setup, short time
reproducibility precision across different operators, instruments, labs or days
resolution the smallest change the instrument can show (one count) a higher-resolution instrument; it does not make the reading more accurate

The GUM method in five steps

  1. Model the result as a function of the inputs, for example P=V2/RP = V^2/R.
  2. Find a standard uncertainty uu (one standard deviation) for every input.
    • Type A (from statistics of repeated readings): nn readings with sample standard deviation ss give uA=s/nu_A = s/\sqrt{n} for their mean.
    • Type B (from any other information: specifications, calibration certificates, resolution). If all you know is that the value lies within ±a\pm a with no value more likely than another, use a rectangular distribution: u=a/3u = a/\sqrt{3}. A display resolution δ\delta means a=δ/2a = \delta/2.
  3. Combine: uc2=∑i(ciui)2u_c^2 = \sum_i (c_i u_i)^2 with sensitivity coefficients ci=∂f/∂xic_i = \partial f/\partial x_i (inputs assumed independent). For a product of powers f=xaybf = x^a y^b it is simpler in relative terms:
(uc(f)f)2=(a u(x)x)2+(b u(y)y)2\left(\frac{u_c(f)}{f}\right)^2 = \left(a\,\frac{u(x)}{x}\right)^2 + \left(b\,\frac{u(y)}{y}\right)^2
  1. Expand: U=k ucU = k\,u_c with coverage factor k=2k = 2, which gives an interval of roughly 95 %95\,\% coverage when the combined distribution is close to normal.
  2. Report: value, UU, kk, rounded so UU has one or two significant figures and the value ends at the same decimal place.

Worked example: power in a resistor

Given: ten readings of the voltage across a 100 Ω100\ \Omega, ±1 %\pm 1\,\% resistor on the 6 V6\ \text{V} range of an example meter with accuracy ±(0.5 %+3 counts)\pm(0.5\,\% + 3\ \text{counts}) and resolution 0.001 V0.001\ \text{V}: mean Vˉ=4.987 V\bar V = 4.987\ \text{V}, sample standard deviation s=0.0035 Vs = 0.0035\ \text{V}.

Find: PP and its expanded uncertainty (k=2k = 2).

Solve:

uA=0.0035 V10=0.0011 Vaspec=0.005×4.987 V+0.003 V=0.0279 V,uspec=0.0279 V3=0.0161 Vures=0.0005 V3=0.0003 Vuc(V)=0.00112+0.01612+0.00032 V=0.0162 Vu(R)=1 Ω3=0.58 ΩP=(4.987 V)2100 Ω=0.2487 Wuc(P)P=(2×0.01624.987)2+(0.58100)2=0.00652+0.00582=0.0087U(P)=2×0.0087×0.2487 W=0.0043 W\begin{aligned} u_A &= \frac{0.0035\ \text{V}}{\sqrt{10}} = 0.0011\ \text{V} \\ a_\text{spec} &= 0.005 \times 4.987\ \text{V} + 0.003\ \text{V} = 0.0279\ \text{V}, \quad u_\text{spec} = \frac{0.0279\ \text{V}}{\sqrt{3}} = 0.0161\ \text{V} \\ u_\text{res} &= \frac{0.0005\ \text{V}}{\sqrt{3}} = 0.0003\ \text{V} \\ u_c(V) &= \sqrt{0.0011^2 + 0.0161^2 + 0.0003^2}\ \text{V} = 0.0162\ \text{V} \\ u(R) &= \frac{1\ \Omega}{\sqrt{3}} = 0.58\ \Omega \\ P &= \frac{(4.987\ \text{V})^2}{100\ \Omega} = 0.2487\ \text{W} \\ \frac{u_c(P)}{P} &= \sqrt{\left(2 \times \frac{0.0162}{4.987}\right)^2 + \left(\frac{0.58}{100}\right)^2} = \sqrt{0.0065^2 + 0.0058^2} = 0.0087 \\ U(P) &= 2 \times 0.0087 \times 0.2487\ \text{W} = 0.0043\ \text{W} \end{aligned}

Result: P=0.249±0.004 WP = 0.249 \pm 0.004\ \text{W} (k=2k = 2).

Check: the meter specification dominates the voltage budget; averaging more readings would shrink uAu_A but barely change the answer. The voltage enters squared, so its relative uncertainty counts twice. A Monte Carlo simulation (sampling every input from its distribution, GUM Supplement 1, JCGM 101:2008) should give the same uc(P)u_c(P):

python · try it yourself
import numpy as np

readings = np.array([4.983, 4.991, 4.985, 4.990, 4.982, 4.988, 4.992, 4.984, 4.989, 4.986])  # volts
n = len(readings)
V = float(readings.mean())
s = float(readings.std(ddof=1))
u_A = s / np.sqrt(n)                          # Type A: standard uncertainty of the mean
a_spec = 0.005 * V + 3 * 0.001                # Type B: example meter spec, half-width in volts
u_spec = a_spec / np.sqrt(3)                  # rectangular distribution
u_res = (0.001 / 2) / np.sqrt(3)              # display resolution
u_V = np.sqrt(u_A**2 + u_spec**2 + u_res**2)
R, u_R = 100.0, 1.0 / np.sqrt(3)              # 100 ohm +/- 1 %, rectangular
P = V**2 / R
u_P = P * np.sqrt((2 * u_V / V)**2 + (u_R / R)**2)
print(f"V = {V:.4f} V, s = {s:.4f} V, u_A = {u_A:.4f} V, u_spec = {u_spec:.4f} V, u_res = {u_res:.4f} V")
print(f"u_c(V) = {u_V:.4f} V")
print(f"P = {P:.4f} W, u_c(P) = {u_P:.4f} W, U(k=2) = {2 * u_P:.4f} W")

rng = np.random.default_rng(1)
N = 200_000
V_mc = V + rng.normal(0, u_A, N) + rng.uniform(-a_spec, a_spec, N) + rng.uniform(-0.0005, 0.0005, N)
R_mc = rng.uniform(99.0, 101.0, N)
P_mc = V_mc**2 / R_mc
print(f"Monte Carlo: P = {P_mc.mean():.4f} W, u(P) = {P_mc.std():.4f} W")
V = 4.9870 V, s = 0.0035 V, u_A = 0.0011 V, u_spec = 0.0161 V, u_res = 0.0003 V
u_c(V) = 0.0162 V
P = 0.2487 W, u_c(P) = 0.0022 W, U(k=2) = 0.0043 W
Monte Carlo: P = 0.2487 W, u(P) = 0.0022 W

Guided lab: repeated readings and a fitted resistance

Hand calculation. For the school "half the range" rule, ten readings that span 0.8 mm0.8\ \text{mm} give Δx=0.4 mm\Delta x = 0.4\ \text{mm}. The GUM Type A value for the same data is s/10s/\sqrt{10}, about five times smaller for a normal spread of ten readings, because it describes the uncertainty of the mean, not of one reading.

Simulation 1. Ten readings of a rod on a vernier caliper, scattered by 0.3 mm0.3\ \text{mm}. With these settings the lab ends at 52.6±0.4 mm52.6 \pm 0.4\ \text{mm} by the half-range rule. Watch how the mean settles as readings arrive:

10 of 10 readings of the length (mm)51.852.052.252.452.652.853.053.2range 0.8 mmx̄ = 52.570result: 52.6 ± 0.4 mmΔx = (max − min)/2, at least ½ division56cm0510main scale 52 mm + vernier line 5 lines up (5 × 0.1 mm)reading 52.5 mm ± 0.05 mm (half of 0.1 mm)

Simulation 2. A resistance from a fitted VV–II line (the lab writes UU for voltage). Six points around a true 10 Ω10\ \Omega with ±4 %\pm 4\,\% error bars: the least-squares gradient is 9.539 Ω9.539\ \Omega, and the steepest and shallowest lines give 9.5±1.0 Ω9.5 \pm 1.0\ \Omega. Try other data sets: the true value stays inside the interval most of the time, which is what an uncertainty promises.

00.10.20.30.40.50.60.702468I (A)U (V)your lineleast-squares fitsteepest / shallowestyour gradient: 10.00 Ωyour Σr² = 0.0907best gradient: 9.539 Ωintercept: 0.172 Vbest Σr² = 0.0536 (least)steepest: 10.56 Ωshallowest: 8.64 ΩΔm = half the gap = 0.96resistance R:9.5 ± 1.0 Ωerror bars ±0.24 V

What you would measure. Set a bench supply to about 5 V5\ \text{V} (current limit 100 mA100\ \text{mA}) across a 100 Ω100\ \Omega, 1 %1\,\%, 0.6 W0.6\ \text{W} metal-film resistor and take ten voltage readings over a minute. Expect them to agree within a few millivolts (repeatability), a mean near 5 V5\ \text{V}, and a budget like the worked example: uc(V)u_c(V) near 0.016 V0.016\ \text{V}, P=0.25±0.004 WP = 0.25 \pm 0.004\ \text{W}. If a second meter disagrees by more than both expanded uncertainties combined, one of them is outside its specification or a systematic effect (loading, a bad lead) is present.

Hazards and safe setup.

  • 5 V5\ \text{V} DC from a current-limited supply; no mains.
  • The resistor dissipates 0.25 W0.25\ \text{W} and gets warm; let it settle before reading (its value drifts with temperature, another input to your budget).
  • Leads in the V jack, not A.

Pitfalls

  • Quoting the resolution as the uncertainty.
  • Using ss instead of s/ns/\sqrt{n} for the uncertainty of a mean (or the reverse, when the question is about a single future reading).
  • Adding uncertainties linearly instead of in quadrature (overestimates) or forgetting the sensitivity coefficient (the factor 2 for V2V^2).
  • Treating a manufacturer's limit ±a\pm a as a standard deviation (u=a/3u = a/\sqrt{3} for a rectangular limit).
  • Too many digits: 0.2487±0.00432 W0.2487 \pm 0.00432\ \text{W} is not more accurate than 0.249±0.004 W0.249 \pm 0.004\ \text{W}.

Practice

  1. A spec says ±0.02 V\pm 0.02\ \text{V}. What standard uncertainty do you assign, and why?
  2. I=V/RI = V/R with u(V)/V=0.4 %u(V)/V = 0.4\,\% and u(R)/R=0.3 %u(R)/R = 0.3\,\%. What is u(I)/Iu(I)/I?
  3. Twenty readings have s=2.0 mVs = 2.0\ \text{mV}. What is uAu_A of their mean?
Answer
  1. 0.02/3=0.012 V0.02/\sqrt{3} = 0.012\ \text{V}: a limit with no other information is modelled as a rectangular distribution.
  2. 0.42+0.32=0.5 %\sqrt{0.4^2 + 0.3^2} = 0.5\,\% (both exponents have magnitude 1).
  3. 2.0/20=0.45 mV2.0/\sqrt{20} = 0.45\ \text{mV}.

Test yourself

A meter shows six digits but is out of calibration by 2 %. Its readings are

  1. accurate and high resolution
  2. high resolution, not accurate
  3. precise, so accurate
  4. Type A uncertain

The standard uncertainty for a rectangular limit ±a\pm a is

  1. aa
  2. a/2a/2
  3. a/3a/\sqrt{3}
  4. 2a2a

For P=V2/RP = V^2/R, a 0.3 %0.3\,\% relative uncertainty in VV contributes

  1. 0.3 %0.3\,\%
  2. 0.6 %0.6\,\%
  3. 0.09 %0.09\,\%
  4. 0.15 %0.15\,\%

What does the expanded uncertainty with k=2k = 2 give?

  1. The maximum possible error
  2. An interval of roughly 95 % coverage when the distribution is near normal
  3. Two standard deviations of a single reading only
  4. The instrument's resolution