AC Bridges: Maxwell, Schering and Wien
An inductor is not just an inductance: its winding has resistance, so it is . A capacitor's dielectric absorbs a little energy each cycle, so it has a loss angle. Measuring the two parts separately, and accurately, was the job of AC bridges for a century, and their balance equations still define how LCR meters, impedance analysers and high-voltage insulation test sets describe components. This lesson generalises the Wheatstone bridge to impedances and works through the three classic circuits.
Phasors and impedance are from Circuit Theory.
The general balance condition
Replace the four resistors by impedances: from the source node a to b, from b to ground, from a to c, from c to ground, with a detector between b and c. The detector reads zero when the two dividers give the same phasor:
This is one complex equation, i.e. two real conditions: the magnitudes and the phases must both match,
so an AC bridge needs two adjustable elements, and the operator alternates between them until the null is deep. A good bridge design makes the two adjustments independent (each one controls one unknown) and the balance frequency-independent, so harmonics in the source do not spoil the null.
Maxwell (Maxwell–Wien) bridge: inductance
The unknown inductor sits in arm 4; arms 2 and 3 are resistors; arm 1 is in parallel with a standard capacitor . With , the balance gives
Both results are free of : the balance holds at every frequency. sets and sets , so the adjustments are independent. Using a capacitor as the standard is the trick: precision capacitors are far closer to ideal than any reference inductor. The quality factor is ; for high- coils becomes impractically large and the Hay bridge (series ) is used instead.
Worked example: a 10 mH coil
Given: a Maxwell bridge at with , balanced at and .
Find: , and at .
Solve:
Check: units: . A small coil with of copper and at is plausible for a few hundred turns of thin wire. ngspice confirms the null and shows how far off a error in either adjustment puts the detector, for of drive:
cat > maxwell.cir <<'EOF'
* Maxwell inductance bridge at 1 kHz: unknown 10 mH with 25 ohm series resistance
V1 a 0 DC 0 AC 1
R1 a b 40k
C1 a b 10n
R2 b 0 1k
R3 a c 1k
Rx c x 25
Lx x 0 10m
.control
ac lin 1 1k 1k
let vd = mag(v(b) - v(c))
echo "balanced (R1 = 40k, C1 = 10n):"
if vd lt 1e-12
echo "vd below 1e-12 V: null"
end
alter C1 = 11n
ac lin 1 1k 1k
let vd = mag(v(b) - v(c))
echo "C1 = 11n:"
print vd
alter C1 = 10n
alter R1 = 30k
ac lin 1 1k 1k
let vd = mag(v(b) - v(c))
echo "R1 = 30k:"
print vd
.endc
.end
EOF
ngspice -b maxwell.cir 2>&1 | grep -E '^(vd|balanced|C1|R1)'balanced (R1 = 40k, C1 = 10n):
vd below 1e-12 V: null
C1 = 11n:
vd = 5.955698e-03
R1 = 30k:
vd = 7.838603e-03
Simulation. The lab solves the same bridge for AC. E1 copies to node d. At the starting value node d reads at and more at higher frequencies, because the term is what is off. Slide to : d drops to picovolts across the whole frequency axis (the balance is frequency-independent; the tiny residual is the solver's gmin conductance).
Schering bridge: capacitance and loss
The unknown capacitor is modelled as in series with a loss resistance ; its dissipation factor is . In the Schering bridge the unknown is in arm 1 (a–b), in arm 2 (b–ground), a lossless standard capacitor in arm 3 (a–c) and in arm 4 (c–ground). Balance gives
sets , sets directly. Because the standard capacitor and the unknown take the high-voltage side and the adjustable arms sit near ground, the Schering bridge became the standard method for dielectric loss of insulation at high voltage (cables, bushings, transformers). That use is professional high-voltage work and is a concept only here; the same bridge works at a few volts on the bench.
Wien bridge: a frequency-selective null
With and in series in arm 1 and in arm 2, and a resistive ratio on the other side, the bridge balances at one frequency only:
The Wien bridge measures frequency (adjust or for a null), notches out the fundamental in classic distortion analysers (what remains is the harmonics and noise), and, with an amplifier, becomes the Wien-bridge oscillator.
import numpy as np
def detector(Za_b, Zb_0, Za_c, Zc_0, V=1.0):
"""Bridge fed with V at node a: |v_b - v_c|."""
return abs(V * Zb_0 / (Za_b + Zb_0) - V * Zc_0 / (Za_c + Zc_0))
par = lambda Z1, Z2: Z1 * Z2 / (Z1 + Z2)
f = 1000.0
w = 2 * np.pi * f
# Maxwell bridge: arm a-b = R1 || C1, b-0 = R2, a-c = R3, c-0 = unknown Rx + j w Lx
R2 = R3 = 1e3
R1, C1 = 40e3, 10e-9 # settings at balance
Lx, Rx = R2 * R3 * C1, R2 * R3 / R1
print(f"Maxwell: Lx = R2 R3 C1 = {1e3 * Lx:.3f} mH, Rx = R2 R3 / R1 = {Rx:.2f} ohm, Q(1 kHz) = {w * Lx / Rx:.3f}")
for label, r1, c1 in (("C1 = 11 nF", 40e3, 11e-9), ("R1 = 30 kohm", 30e3, 10e-9)):
Z1 = par(r1, 1 / (1j * w * c1))
print(f" off balance, {label}: |vd| = {1e3 * detector(Z1, R2, R3, 25 + 1j * w * 10e-3):.4f} mV per volt")
# Schering bridge: arm a-b = unknown (Cx + Rx series), b-0 = R3, a-c = C2 (standard), c-0 = R4 || C4
C2, R3s, R4, C4 = 100e-12, 1e3, 4.7e3, 67.7e-12
Cx, Rxs = C2 * R4 / R3s, R3s * C4 / C2
print(f"Schering: Cx = C2 R4 / R3 = {1e12 * Cx:.1f} pF, Rx = R3 C4 / C2 = {Rxs:.0f} ohm, "
f"D = w R4 C4 = {w * R4 * C4:.5f}")
# Wien bridge: a-b = R + C series, b-0 = R || C, a-c = R3, c-0 = R4 with R3 = 2 R4
R, C = 10e3, 15.915e-9
f0 = 1 / (2 * np.pi * R * C)
print(f"Wien: f0 = 1/(2 pi R C) = {f0:.1f} Hz")
for fx in (500.0, 900.0, f0, 1100.0, 2000.0):
wx = 2 * np.pi * fx
vd = detector(R + 1 / (1j * wx * C), par(R, 1 / (1j * wx * C)), 2e3, 1e3)
print(f" f = {fx:7.1f} Hz: |vd| = {1e3 * vd:7.3f} mV per volt")Maxwell: Lx = R2 R3 C1 = 10.000 mH, Rx = R2 R3 / R1 = 25.00 ohm, Q(1 kHz) = 2.513
off balance, C1 = 11 nF: |vd| = 5.9557 mV per volt
off balance, R1 = 30 kohm: |vd| = 7.8386 mV per volt
Schering: Cx = C2 R4 / R3 = 470.0 pF, Rx = R3 C4 / C2 = 677 ohm, D = w R4 C4 = 0.00200
Wien: f0 = 1/(2 pi R C) = 1000.0 Hz
f = 500.0 Hz: |vd| = 149.077 mV per volt
f = 900.0 Hz: |vd| = 23.406 mV per volt
f = 1000.0 Hz: |vd| = 0.000 mV per volt
f = 1100.0 Hz: |vd| = 21.162 mV per volt
f = 2000.0 Hz: |vd| = 149.065 mV per volt
The off-balance Maxwell values match ngspice to all printed digits. The Wien detector falls to zero only at ; a frequency error leaves about of the drive, which is why the null is a sensitive frequency indicator.
Guided lab: balance a Maxwell bridge on the bench
Hand calculation. For a inductor (check its datasheet DC resistance, say to ) and : and to .
Simulation. Use the lab with your inductor's values; the lab's slider is only, so account for a different coil resistance through (a larger needs a smaller ).
What you would measure. Drive the bridge from a function generator at , ; use a capacitance decade (or switched , , capacitors) for and a potentiometer in series with for . Watch the detector with a scope in differential mode (channel 1 minus channel 2, both probes 10x) or with a battery-powered DMM on AC volts. Expect the null to fall from tens of millivolts to below a millivolt, limited by stray capacitance and the generator's harmonics; your should be within the inductor's tolerance (often ).
Hazards and safe setup. Generator signal levels only. The detector is floating: do not connect a grounded scope probe's clip to node b or c, it would short an arm through earth.
Pitfalls
- Adjusting only one element and expecting a null: two conditions need two adjustments.
- Ignoring stray capacitance to ground at high frequencies (a Wagner earth or a shielded transformer ratio arm is the classic cure).
- Grounding the detector.
- Reading a Schering balance as "capacitance only": the second adjustment is the loss, often the quantity you need.
Practice
- A Maxwell bridge with balances at , . Find and .
- A Wien bridge uses and . Where is its null?
Answer
- ; .
- .
Test yourself
Why does an AC bridge need two adjustable elements?
- one for each supply terminal
- the complex balance equation is two real conditions
- to double the sensitivity
- to cancel thermal EMFs
In a Maxwell bridge the unknown inductance is measured against
- a standard inductor
- a standard capacitor
- a frequency standard
- a Zener reference
The Schering bridge is best known for measuring
- inductance of coils
- capacitance and dielectric loss (tan delta)
- frequency
- low resistance
A Wien bridge with equal R and C balances at
- every frequency