True RMS, Average-Responding Meters and Crest Factor

A technician checks the current drawn by a small switch-mode power supply (on its low-voltage DC output side, or in a simulation of its mains input) with an inexpensive clamp meter and an expensive one. The readings differ by a factor of two. Both meters are within calibration. The cheap one is average-responding: it measures the rectified mean and multiplies by the factor that is correct for a sine wave. The current is not a sine wave. This lesson explains what each kind of AC meter measures, how large the errors get, and what the crest factor specification protects you from.

EE Foundations defined RMS; Signals and Fourier shows why the RMS of a sum of harmonics is the root-sum-square of their RMS values.

Three numbers of a periodic waveform

For a waveform x(t)x(t) of period TT:

Xrms=1T∫0Tx2 dt,Xavg=1T∫0T∣x∣ dt,Xpk=max⁡∣x∣X_\text{rms} = \sqrt{\frac{1}{T}\int_0^T x^2\,dt}, \qquad X_\text{avg} = \frac{1}{T}\int_0^T \lvert x \rvert\,dt, \qquad X_\text{pk} = \max \lvert x \rvert

and two ratios that characterise its shape:

form factor F=XrmsXavg,crest factor C=XpkXrms\text{form factor } F = \frac{X_\text{rms}}{X_\text{avg}}, \qquad \text{crest factor } C = \frac{X_\text{pk}}{X_\text{rms}}
peak Xpk RMS = 0.707 Xpk average |x| = 0.637 Xpk rectified (dashed) form factor F = 1.111 crest factor C = 1.414 t

For a sine, Xrms=Xpk/2X_\text{rms} = X_\text{pk}/\sqrt{2}, Xavg=2Xpk/πX_\text{avg} = 2X_\text{pk}/\pi, so F=π/(22)=1.1107F = \pi/(2\sqrt{2}) = 1.1107 and C=2=1.414C = \sqrt{2} = 1.414. A square wave has F=C=1F = C = 1; a triangle F=1.155F = 1.155, C=1.732C = 1.732.

How AC meters respond

Type Measures Correct for Notes
average-responding, RMS-calibrated 1.1107×1.1107 \times rectified mean sines only cheap meters, older clamps; error +11 %+11\,\% on a square, large and negative on peaky waveforms
peak-responding Xpk/2X_\text{pk}/\sqrt{2} sines only some RF probes and simple AC adapters
true RMS, analog converter x2‾\sqrt{\overline{x^2}} by a log/antilog or thermal circuit any waveform within bandwidth and crest factor the classic RMS-to-DC converter IC; needs a minimum signal (a few percent of range) to settle
true RMS, sampling sampled xkx_k, computed 1N∑xk2\sqrt{\frac{1}{N}\sum x_k^2} any waveform below the sample rate's limits most modern bench DMMs and power analysers
thermal converter heating of a resistor compared with DC any waveform, very wide bandwidth AC–DC transfer standards in calibration labs

Most meters measure the RMS of the AC component (input capacitor-coupled) on their AC function. The total RMS of a signal with a DC part is

Xrms, total=XDC2+Xrms, AC2X_\text{rms, total} = \sqrt{X_\text{DC}^2 + X_\text{rms, AC}^2}

so for a unipolar pulse train you must combine the DC and AC readings (some meters offer an "AC+DC" function that does it).

Crest factor and bandwidth

A true RMS converter only computes the right value if its front end passes the peaks without clipping and its bandwidth covers the significant harmonics. Datasheets therefore state a maximum crest factor (for example 33 at full scale, more at lower readings) and often an additional error that grows with crest factor (illustrative figures: +0.05 %+0.05\,\% for CC from 1 to 2, +0.2 %+0.2\,\% from 2 to 3; check your meter). A narrow pulse train has a huge crest factor: for a pulse of duty DD, C=(1−D)/DC = \sqrt{(1 - D)/D} for the AC component, 33 at D=10 %D = 10\,\% and 77 at D=2 %D = 2\,\%. Its harmonics also extend far beyond the fundamental, so a meter with a 100 kHz100\ \text{kHz} AC bandwidth will read low on a 20 kHz20\ \text{kHz} pulse train with fast edges.

Worked example: a 10 percent PWM signal

Given: a 00 to 5 V5\ \text{V} PWM signal with duty D=0.10D = 0.10.

Find: the DC reading, the true RMS of the AC component, the total RMS, the crest factor of the AC component, and what an average-responding AC meter shows.

Solve:

VDC=D Vpk=0.5 V,Vrms, total=VpkD=50.1 V=1.581 VVrms, AC=1.5812−0.52 V=2.5−0.25 V=1.500 VCAC=5−0.51.500=3.0Vavg, AC=2D(1−D) Vpk=0.9 V,reading=1.1107×0.9 V=1.000 V\begin{aligned} V_\text{DC} &= D\,V_\text{pk} = 0.5\ \text{V}, \qquad V_\text{rms, total} = V_\text{pk}\sqrt{D} = 5\sqrt{0.1}\ \text{V} = 1.581\ \text{V} \\ V_\text{rms, AC} &= \sqrt{1.581^2 - 0.5^2}\ \text{V} = \sqrt{2.5 - 0.25}\ \text{V} = 1.500\ \text{V} \\ C_\text{AC} &= \frac{5 - 0.5}{1.500} = 3.0 \\ V_\text{avg, AC} &= 2D(1 - D)\,V_\text{pk} = 0.9\ \text{V}, \qquad \text{reading} = 1.1107 \times 0.9\ \text{V} = 1.000\ \text{V} \end{aligned}

Check: the average-responding meter reads 1.000 V1.000\ \text{V}, 33 %33\,\% below the true 1.500 V1.500\ \text{V} of the AC component. A crest factor of 33 sits exactly at a typical meter's full-scale limit: measure on a range where the reading is at most half of full scale to leave headroom. Units are volts throughout.

python · try it yourself
import numpy as np

N = 200_000
t = np.arange(N) / N                       # one period, normalised time 0..1
FF_SINE = np.pi / (2 * np.sqrt(2))         # form factor of a sine, 1.1107

def pulse_current(t, width=0.125):         # cos^2 current pulses at the voltage peaks, alternating sign
    x = np.zeros_like(t)
    for centre, sign in ((0.25, 1), (0.75, -1)):
        m = np.abs(t - centre) < width / 2
        x[m] = sign * np.cos(np.pi * (t[m] - centre) / width) ** 2
    return x

waves = {
    "sine":            np.sin(2 * np.pi * t),
    "square":          np.where(t < 0.5, 1.0, -1.0),
    "triangle":        2 * np.abs(2 * (t - np.floor(t + 0.5))) - 1,
    "half-wave sine":  np.maximum(np.sin(2 * np.pi * t), 0.0),
    "PWM 10 %, 0/5 V": np.where(t < 0.1, 5.0, 0.0),
    "rectifier input current": pulse_current(t),
}
print(f"{'waveform':24s} {'avg-resp':>8s} {'TRMS ac':>8s} {'TRMS ac+dc':>10s} {'error':>7s} {'CF(ac)':>6s}")
for name, x in waves.items():
    ac = x - x.mean()                      # AC coupling removes the DC component
    avg_resp = FF_SINE * np.mean(np.abs(ac))
    rms_ac = np.sqrt(np.mean(ac**2))
    rms_total = np.sqrt(np.mean(x**2))
    err = round(100 * (avg_resp / rms_ac - 1), 1) + 0.0   # + 0.0 turns -0.0 into 0.0
    cf = np.max(np.abs(ac)) / rms_ac
    print(f"{name:24s} {avg_resp:8.4f} {rms_ac:8.4f} {rms_total:10.4f} {err:+6.1f}% {cf:6.2f}")
waveform                 avg-resp  TRMS ac TRMS ac+dc   error CF(ac)
sine                       0.7071   0.7071     0.7071   +0.0%   1.41
square                     1.1107   1.0000     1.0000  +11.1%   1.00
triangle                   0.5554   0.5774     0.5774   -3.8%   1.73
half-wave sine             0.3897   0.3856     0.5000   +1.1%   1.77
PWM 10 %, 0/5 V            0.9996   1.5000     1.5811  -33.4%   3.00
rectifier input current    0.1388   0.3062     0.3062  -54.7%   3.27

The last row is the classic case of the opening: the current drawn by a rectifier with a smoothing capacitor flows in short pulses near the voltage peaks. An average-responding meter reads it 55 %55\,\% low. (This current waveform is computed here; mains-input currents are measured only by trained people with rated equipment, Power Measurement.)

Guided lab: RMS of a sine and of a square from a generator

Hand calculation. A sine of amplitude U0=10 VU_0 = 10\ \text{V} has Urms=7.07 VU_\text{rms} = 7.07\ \text{V} and delivers U02/(2R)=5 WU_0^2/(2R) = 5\ \text{W} into 10 Ω10\ \Omega; a square of the same amplitude has Urms=10 VU_\text{rms} = 10\ \text{V} and delivers 10 W10\ \text{W}.

Simulation. The AC lab (it writes voltage as UU) shows the sine, the ±Urms\pm U_\text{rms} lines and the average power: at these settings Urms=7.07 VU_\text{rms} = 7.07\ \text{V} and P=5 WP = 5\ \text{W}. Use only the "your own" preset here.

U0rotating phasoru (V)tUrms = 7.07 V10.0−10.0T = 2.00 sp = u²/R (W)average P = 5.00 W10.00R = 10 Ωi = 0.00 A

What you would measure. Set a function generator to 1 kHz1\ \text{kHz}, 2 Vpp2\ \text{V}_\text{pp} into a high-impedance load and read it with a true RMS DMM and, if you have one, an average-responding meter. Expected (true RMS / average-responding): sine 0.707/0.707 V0.707 / 0.707\ \text{V}, square 1.000/1.111 V1.000 / 1.111\ \text{V}, triangle 0.577/0.555 V0.577 / 0.555\ \text{V}, each within the meters' AC specification (often ±0.5\pm 0.5 to 1 %1\,\% at 1 kHz1\ \text{kHz}). Then try a 10 %10\,\% pulse, 00 to 2 V2\ \text{V}: true RMS AC about 0.600 V0.600\ \text{V}, DC 0.200 V0.200\ \text{V}, average-responding about 0.400 V0.400\ \text{V}. Raise the frequency to 50 kHz50\ \text{kHz} and watch the true RMS reading fall as harmonics leave the meter's bandwidth.

Hazards and safe setup. Signal-generator levels only (≤10 Vpp\le 10\ \text{V}_\text{pp}). Never "check" mains RMS with a bench DMM's leads as an exercise in this course: mains measurements need CAT-rated meters and training.

Pitfalls

  • Trusting an average-responding meter on anything but a clean sine.
  • Reading only the AC function of a unipolar signal and calling it the RMS.
  • Ignoring crest-factor limits and bandwidth: a true RMS meter is only true within them.
  • Measuring small AC signals on a true RMS meter below its minimum input (often 11 to 5 %5\,\% of range), where the converter's own noise dominates.

Practice

  1. A square wave of ±3 V\pm 3\ \text{V} is read by an average-responding meter. What does it show?
  2. What crest factor does a 00/11 pulse train with D=0.04D = 0.04 have for its AC component?
Answer
  1. Xavg=3 VX_\text{avg} = 3\ \text{V}, reading 1.1107×3=3.33 V1.1107 \times 3 = 3.33\ \text{V}; the true RMS is 3.00 V3.00\ \text{V}.
  2. C=0.96/0.04=24=4.9C = \sqrt{0.96/0.04} = \sqrt{24} = 4.9.

Test yourself

An average-responding meter reads a square wave

  1. correctly
  2. about 11 % high
  3. about 11 % low
  4. as zero

The crest factor is

  1. RMS divided by average
  2. peak divided by RMS
  3. peak divided by average
  4. peak-to-peak divided by peak

A meter's AC function reads 1.50 V and its DC function 0.50 V on the same signal. The total RMS is

  1. 2.00 V
  2. 1.00 V
  3. 1.58 V
  4. 1.50 V

Why can a true RMS meter read a fast pulse train low?

  1. its crest factor and bandwidth limits
  2. it averages instead
  3. pulses have no RMS
  4. the DC component is too large