Probes: Compensation, Loading, Differential and Current Probes
The probe is the part of the oscilloscope that touches your circuit, and the part most often blamed last. A mis-adjusted 10x probe turns a square wave into a shark fin; a 1x probe on a node cuts a signal to about a sixth; a long ground clip makes a clean edge ring at ; and a grounded probe clipped to the wrong node shorts it to earth. This lesson models the passive probe as the circuit it is, quantifies its loading, and introduces the probes for jobs it cannot do: active, differential and current probes. EE Foundations introduced 1x and 10x probes and the compensation trimmer.
The 10x passive probe is a compensated divider
The tip contains in parallel with a small capacitor . The scope's input is in parallel with , which includes the input capacitance (about ) and the cable's (often to ). At DC the divider is ; at high frequency it is the capacitive divider . The two agree at every frequency when
which is the compensation condition: the trimmer (in the tip or the connector) adjusts the effective capacitance until a square wave from the scope's calibrator is flat. If is too large (over-compensated) the edge jumps above its final value and decays; too small, it rises in two stages. The tail's time constant is
so mis-compensation shows at kilohertz frequencies as a slanted top, and gives amplitude errors of tens of percent at all frequencies above about . Compensate every probe on the channel you will use it on.
Simulation. The netlist is the probe and scope input driven by a step from a generator. At the default the output at node out jumps to and decays to with ; set for a flat top, and for the slow rise.
The same exponential tail, scaled up a millionfold in time, is the RC decay in the transient lab: with and its is also , and the "switch off" curve is the shape of an over-compensated probe's overshoot dying away (here from ; in the probe from above the final value).
Loading: the probe is part of the circuit
A compensated 10x probe presents in parallel with . At DC the is the loading error you already know; at high frequency the capacitance dominates: is at and at . A 1x probe connects the cable directly: in parallel with about , ten times worse, and its bandwidth is often only to .
Worked example: probing a 1 kilohm node
Given: a node with a Thevenin resistance of ; a 1x probe (with scope: ) and a 10x probe ().
Find: the bandwidth the probe leaves at that node.
Solve: the source resistance and the probe capacitance form a low-pass filter:
Check: whatever the scope's bandwidth, a 1x probe on this node cannot show more than about . Lower the source impedance (probe a buffered node) or use an active probe. ngspice gives the same with the exact loads:
cat > probe.cir <<'EOF'
* A 1 kohm source node probed with a 1x and a 10x passive probe (tip models R || C)
V1 s 0 DC 0 AC 1
Rs s n 1k
* 1x probe + scope input: 1 Mohm || 100 pF (cable capacitance dominates)
Rp1 n 0 1meg
Cp1 n 0 100p
.control
ac dec 50 10k 100meg
meas ac f3db_1x when vdb(n)=-3 fall=1
* 10x probe: 10 Mohm || 10 pF at the tip
alter Rp1 = 10meg
alter Cp1 = 10p
ac dec 50 10k 100meg
meas ac f3db_10x when vdb(n)=-3 fall=1
.endc
.end
EOF
ngspice -b probe.cir 2>&1 | grep -E '^f3db'f3db_1x = 1.586178e+06
f3db_10x = 1.587616e+07
The program below summarises compensation, loading and the ground lead:
import numpy as np
# 10x passive probe: tip 9 Mohm || C1 into scope 1 Mohm || C2 (input 15 pF + cable 85 pF)
R1, R2, C2 = 9e6, 1e6, 100e-12
C1 = R2 * C2 / R1 # compensation: R1 C1 = R2 C2
C_tip = C1 * C2 / (C1 + C2) # capacitance the circuit sees
print(f"compensated C1 = {1e12 * C1:.2f} pF, tip loading 10 Mohm || {1e12 * C_tip:.1f} pF")
for f in (1e3, 1e6, 10e6, 100e6):
Z = 1 / (1 / (R1 + R2) + 2j * np.pi * f * C_tip)
print(f" at {f / 1e6:7.3f} MHz: |Z_tip| = {abs(Z) / 1e3:9.2f} kohm")
# Mis-compensation: initial step value C1/(C1+C2) against the DC value R2/(R1+R2), tail time constant
for C1x in (8e-12, 11.11e-12, 15e-12):
v0, vdc = C1x / (C1x + C2), R2 / (R1 + R2)
tau = (R1 * R2 / (R1 + R2)) * (C1x + C2)
print(f"C1 = {1e12 * C1x:5.2f} pF: edge jumps to {100 * v0 / vdc:5.1f} % of final, settles with tau = {1e6 * tau:5.1f} us")
# Ground-lead inductance resonating with the tip capacitance
for name, L in (("15 cm ground clip", 150e-9), ("spring tip, 1 cm", 10e-9)):
f_ring = 1 / (2 * np.pi * np.sqrt(L * C_tip))
print(f"{name:17s}: L = {1e9 * L:5.0f} nH -> rings at {f_ring / 1e6:5.0f} MHz")compensated C1 = 11.11 pF, tip loading 10 Mohm || 10.0 pF
at 0.001 MHz: |Z_tip| = 8467.33 kohm
at 1.000 MHz: |Z_tip| = 15.92 kohm
at 10.000 MHz: |Z_tip| = 1.59 kohm
at 100.000 MHz: |Z_tip| = 0.16 kohm
C1 = 8.00 pF: edge jumps to 74.1 % of final, settles with tau = 97.2 us
C1 = 11.11 pF: edge jumps to 100.0 % of final, settles with tau = 100.0 us
C1 = 15.00 pF: edge jumps to 130.4 % of final, settles with tau = 103.5 us
15 cm ground clip: L = 150 nH -> rings at 130 MHz
spring tip, 1 cm : L = 10 nH -> rings at 503 MHz
The ground lead is an inductor
The probe's ground clip is a loop of wire, about per millimetre: a clip is roughly . With the tip capacitance it forms a series resonant circuit that rings on every fast edge (here at about ), and the loop picks up magnetic fields from nearby switching currents. For fast edges use the short spring ground or a probe socket on the board (), which moves the resonance above the scope's bandwidth.
Beyond the passive probe
- Active (FET) probes put a buffer amplifier in the tip: about and , bandwidths of GHz, limited input range (a few volts). For fast logic and high-impedance nodes.
- Differential probes measure with high common-mode rejection (CMRR), so neither input needs to be at ground: high-side shunts, gate drives, motor phases, the floating detector of a bridge. High-voltage differential probes are the safe way to view mains-referenced signals; they are used by trained people on rated equipment. Never "float" a scope by defeating its protective earth to make a measurement: the whole chassis then sits at the circuit's voltage.
- Current probes clamp around a wire: AC-only transformer probes, or AC/DC probes with a Hall element and a compensating winding (Current Measurement). Degauss and zero them before use; they add a small insertion impedance.
Every passive probe's ground clip is connected to the scope chassis and, through the mains cord, to protective earth. Clip it only to a point that is at (or meant to be at) earth potential, such as the circuit ground of a bench-supplied circuit.
Guided lab: compensate a probe and measure its loading
Hand calculation. A 10x probe on a scope with a calibrator: a flat top when compensated; with the trimmer far off, an overshoot or undershoot of to decaying with . On a source ( series resistor from a generator), the 10x probe's gives , the 1x probe's .
Simulation. The two labs and the ngspice deck with Rs s n 10k.
What you would measure. Connect the probe to the calibrator, set and adjust the trimmer with an insulated tool until the top is flat. Then drive a resistor from a generator ( sine, sweep to ) and watch the amplitude at the resistor's far end fall to near with the 10x probe and near with 1x (within about , depending on your cable). Finally, view a fast generator edge with the long ground clip and with a spring tip: the ringing should mostly disappear.
Hazards and safe setup. Generator and calibrator levels only. The ground clip goes to circuit ground; no mains-referenced circuits.
Pitfalls
- Using a probe compensated on another scope or channel.
- Setting the scope channel to 1x with a 10x probe (readings ten times too small) or the reverse.
- Measuring fast edges with a long ground lead and blaming the circuit for the ringing.
- Clipping a grounded probe to a non-ground node; use a differential probe or two channels in math mode (A − B) for small signals only.
Practice
- A scope's input is and the cable adds . What compensates a tip?
- A probe with tip capacitance and a ground lead: at what frequency does it ring?
Answer
- , .
- .
Test yourself
A 10x probe is correctly compensated when
- the trimmer is at mid-travel
At 100 MHz the loading of a 10x passive probe is dominated by
- its 10 M ohm resistance
- its tip capacitance, about of reactance
- the ground lead resistance
- the scope's ADC
A square wave shows ringing at about 100 MHz only when the long ground clip is used. The cause is
- the circuit oscillates
- the ground lead inductance resonating with the tip capacitance
- the probe is over-compensated
- aliasing
To view the voltage across a high-side shunt you should use
- a 10x probe with its ground clip on the shunt's low end
- a differential probe
- a 1x probe
- a current probe on the scope's ground