The probe is the part of the oscilloscope that touches your circuit, and the part most often blamed last. A mis-adjusted 10x probe turns a square wave into a shark fin; a 1x probe on a 1 kΩ1\ \text{k}\Omega node cuts a 10 MHz10\ \text{MHz} signal to about a sixth; a long ground clip makes a clean edge ring at 130 MHz130\ \text{MHz}; and a grounded probe clipped to the wrong node shorts it to earth. This lesson models the passive probe as the circuit it is, quantifies its loading, and introduces the probes for jobs it cannot do: active, differential and current probes. EE Foundations introduced 1x and 10x probes and the compensation trimmer.

The 10x passive probe is a compensated divider

tip R1 9 MOhm C1 trimmer probe tip coax cable (part of C2) R2 1 MOhm C2 100 pF scope input ground clip ~150 nH circuit ground compensated when R1 C1 = R2 C2

The tip contains R1=9 MΩR_1 = 9\ \text{M}\Omega in parallel with a small capacitor C1C_1. The scope's input is R2=1 MΩR_2 = 1\ \text{M}\Omega in parallel with C2C_2, which includes the input capacitance (about 15 pF15\ \text{pF}) and the cable's (often 5050 to 100 pF100\ \text{pF}). At DC the divider is R2/(R1+R2)=1/10R_2/(R_1 + R_2) = 1/10; at high frequency it is the capacitive divider C1/(C1+C2)C_1/(C_1 + C_2). The two agree at every frequency when

R1C1=R2C2R_1 C_1 = R_2 C_2

which is the compensation condition: the trimmer (in the tip or the connector) adjusts the effective capacitance until a square wave from the scope's calibrator is flat. If C1C_1 is too large (over-compensated) the edge jumps above its final value and decays; too small, it rises in two stages. The tail's time constant is

τ=(R1∥R2)(C1+C2)≈900 kΩ×111 pF≈100 μs\tau = (R_1 \parallel R_2)(C_1 + C_2) \approx 900\ \text{k}\Omega \times 111\ \text{pF} \approx 100\ \mu\text{s}

so mis-compensation shows at kilohertz frequencies as a slanted top, and gives amplitude errors of tens of percent at all frequencies above about 1/(2πτ)≈1.6 kHz1/(2\pi\tau) \approx 1.6\ \text{kHz}. Compensate every probe on the channel you will use it on.

Simulation. The netlist is the probe and scope input driven by a step from a generator. At the default C1=15 pFC_1 = 15\ \text{pF} the output at node out jumps to 0.130 V0.130\ \text{V} and decays to 0.100 V0.100\ \text{V} with τ≈104 μs\tau \approx 104\ \mu\text{s}; set C1=11.1 pFC_1 = 11.1\ \text{pF} for a flat top, and 8 pF8\ \text{pF} for the slow rise.

0.000.050.100100200300400t (us)VV(out)at t = 400.0 usV(in)1.00 VV(tip)1.00 VV(out)101 mVI(V1)-213 nAI(Rg)213 nAI(R1)99.9 nAI(C1)113 nAI(R2)101 nAI(C2)113 nA

The same exponential tail, scaled up a millionfold in time, is the RC decay in the transient lab: with R=1 kΩR = 1\ \text{k}\Omega and C=0.1 μFC = 0.1\ \mu\text{F} its τ\tau is also 100 μs100\ \mu\text{s}, and the "switch off" curve is the shape of an over-compensated probe's overshoot dying away (here from 10 V10\ \text{V}; in the probe from 30 %30\,\% above the final value).

+-R = 1.00 kOhmsource shorted at t = 0V = 10.0 V, C = 100 nFtau = RC = 100 ust600 ust/tau6.00vC24.8 mVi-24.8 uAvR-24.8 mV-10.0-5.00.05.010.001 tau2 tau3 tau4 tau5 tau6 tautv (V)-10 mA0 A10 mAi36.8 %5 tau: 0.7 %vC(t), capacitor voltagei(t)

Loading: the probe is part of the circuit

A compensated 10x probe presents 10 MΩ10\ \text{M}\Omega in parallel with Ctip=C1C2/(C1+C2)≈10 pFC_\text{tip} = C_1 C_2/(C_1 + C_2) \approx 10\ \text{pF}. At DC the 10 MΩ10\ \text{M}\Omega is the loading error you already know; at high frequency the capacitance dominates: ∣Z∣=1/(2πfC)\lvert Z \rvert = 1/(2\pi f C) is 16 kΩ16\ \text{k}\Omega at 1 MHz1\ \text{MHz} and 160 Ω160\ \Omega at 100 MHz100\ \text{MHz}. A 1x probe connects the cable directly: 1 MΩ1\ \text{M}\Omega in parallel with about 100 pF100\ \text{pF}, ten times worse, and its bandwidth is often only 66 to 10 MHz10\ \text{MHz}.

Worked example: probing a 1 kilohm node

Given: a node with a Thevenin resistance of 1 kΩ1\ \text{k}\Omega; a 1x probe (with scope: 1 MΩ∥100 pF1\ \text{M}\Omega \parallel 100\ \text{pF}) and a 10x probe (10 MΩ∥10 pF10\ \text{M}\Omega \parallel 10\ \text{pF}).

Find: the bandwidth the probe leaves at that node.

Solve: the source resistance and the probe capacitance form a low-pass filter:

f−3 dB≈12πRThCtip:12π×1 kΩ×100 pF=1.59 MHz,12π×1 kΩ×10 pF=15.9 MHzf_{-3\,\text{dB}} \approx \frac{1}{2\pi R_\text{Th} C_\text{tip}}: \quad \frac{1}{2\pi \times 1\ \text{k}\Omega \times 100\ \text{pF}} = 1.59\ \text{MHz}, \qquad \frac{1}{2\pi \times 1\ \text{k}\Omega \times 10\ \text{pF}} = 15.9\ \text{MHz}

Check: whatever the scope's bandwidth, a 1x probe on this node cannot show more than about 1.6 MHz1.6\ \text{MHz}. Lower the source impedance (probe a buffered node) or use an active probe. ngspice gives the same with the exact R∥CR \parallel C loads:

bash · try it yourself
cat > probe.cir <<'EOF'
* A 1 kohm source node probed with a 1x and a 10x passive probe (tip models R || C)
V1 s 0 DC 0 AC 1
Rs s n 1k
* 1x probe + scope input: 1 Mohm || 100 pF (cable capacitance dominates)
Rp1 n 0 1meg
Cp1 n 0 100p
.control
ac dec 50 10k 100meg
meas ac f3db_1x when vdb(n)=-3 fall=1
* 10x probe: 10 Mohm || 10 pF at the tip
alter Rp1 = 10meg
alter Cp1 = 10p
ac dec 50 10k 100meg
meas ac f3db_10x when vdb(n)=-3 fall=1
.endc
.end
EOF
ngspice -b probe.cir 2>&1 | grep -E '^f3db'
f3db_1x             =  1.586178e+06
f3db_10x            =  1.587616e+07

The program below summarises compensation, loading and the ground lead:

python · try it yourself
import numpy as np

# 10x passive probe: tip 9 Mohm || C1 into scope 1 Mohm || C2 (input 15 pF + cable 85 pF)
R1, R2, C2 = 9e6, 1e6, 100e-12
C1 = R2 * C2 / R1                                  # compensation: R1 C1 = R2 C2
C_tip = C1 * C2 / (C1 + C2)                        # capacitance the circuit sees
print(f"compensated C1 = {1e12 * C1:.2f} pF, tip loading 10 Mohm || {1e12 * C_tip:.1f} pF")
for f in (1e3, 1e6, 10e6, 100e6):
    Z = 1 / (1 / (R1 + R2) + 2j * np.pi * f * C_tip)
    print(f"  at {f / 1e6:7.3f} MHz: |Z_tip| = {abs(Z) / 1e3:9.2f} kohm")

# Mis-compensation: initial step value C1/(C1+C2) against the DC value R2/(R1+R2), tail time constant
for C1x in (8e-12, 11.11e-12, 15e-12):
    v0, vdc = C1x / (C1x + C2), R2 / (R1 + R2)
    tau = (R1 * R2 / (R1 + R2)) * (C1x + C2)
    print(f"C1 = {1e12 * C1x:5.2f} pF: edge jumps to {100 * v0 / vdc:5.1f} % of final, settles with tau = {1e6 * tau:5.1f} us")

# Ground-lead inductance resonating with the tip capacitance
for name, L in (("15 cm ground clip", 150e-9), ("spring tip, 1 cm", 10e-9)):
    f_ring = 1 / (2 * np.pi * np.sqrt(L * C_tip))
    print(f"{name:17s}: L = {1e9 * L:5.0f} nH -> rings at {f_ring / 1e6:5.0f} MHz")
compensated C1 = 11.11 pF, tip loading 10 Mohm || 10.0 pF
  at   0.001 MHz: |Z_tip| =   8467.33 kohm
  at   1.000 MHz: |Z_tip| =     15.92 kohm
  at  10.000 MHz: |Z_tip| =      1.59 kohm
  at 100.000 MHz: |Z_tip| =      0.16 kohm
C1 =  8.00 pF: edge jumps to  74.1 % of final, settles with tau =  97.2 us
C1 = 11.11 pF: edge jumps to 100.0 % of final, settles with tau = 100.0 us
C1 = 15.00 pF: edge jumps to 130.4 % of final, settles with tau = 103.5 us
15 cm ground clip: L =   150 nH -> rings at   130 MHz
spring tip, 1 cm : L =    10 nH -> rings at   503 MHz

The ground lead is an inductor

The probe's ground clip is a loop of wire, about 1 nH1\ \text{nH} per millimetre: a 15 cm15\ \text{cm} clip is roughly 150 nH150\ \text{nH}. With the tip capacitance it forms a series resonant circuit that rings on every fast edge (here at about 130 MHz130\ \text{MHz}), and the loop picks up magnetic fields from nearby switching currents. For fast edges use the short spring ground or a probe socket on the board (≤1 cm\le 1\ \text{cm}), which moves the resonance above the scope's bandwidth.

Beyond the passive probe

  • Active (FET) probes put a buffer amplifier in the tip: about 1 pF1\ \text{pF} and 1 MΩ1\ \text{M}\Omega, bandwidths of GHz, limited input range (a few volts). For fast logic and high-impedance nodes.
  • Differential probes measure v+−v−v_+ - v_- with high common-mode rejection (CMRR), so neither input needs to be at ground: high-side shunts, gate drives, motor phases, the floating detector of a bridge. High-voltage differential probes are the safe way to view mains-referenced signals; they are used by trained people on rated equipment. Never "float" a scope by defeating its protective earth to make a measurement: the whole chassis then sits at the circuit's voltage.
  • Current probes clamp around a wire: AC-only transformer probes, or AC/DC probes with a Hall element and a compensating winding (Current Measurement). Degauss and zero them before use; they add a small insertion impedance.

Every passive probe's ground clip is connected to the scope chassis and, through the mains cord, to protective earth. Clip it only to a point that is at (or meant to be at) earth potential, such as the circuit ground of a bench-supplied circuit.

Guided lab: compensate a probe and measure its loading

Hand calculation. A 10x probe on a scope with a 1 kHz1\ \text{kHz} calibrator: a flat top when compensated; with the trimmer far off, an overshoot or undershoot of 1010 to 30 %30\,\% decaying with τ≈100 μs\tau \approx 100\ \mu\text{s}. On a 10 kΩ10\ \text{k}\Omega source (10 kΩ10\ \text{k}\Omega series resistor from a generator), the 10x probe's 10 pF10\ \text{pF} gives f−3 dB=1.6 MHzf_{-3\,\text{dB}} = 1.6\ \text{MHz}, the 1x probe's 100 pF100\ \text{pF} 160 kHz160\ \text{kHz}.

Simulation. The two labs and the ngspice deck with Rs s n 10k.

What you would measure. Connect the probe to the calibrator, set 0.2 ms/div0.2\ \text{ms/div} and adjust the trimmer with an insulated tool until the top is flat. Then drive a 10 kΩ10\ \text{k}\Omega resistor from a generator (1 V1\ \text{V} sine, sweep 10 kHz10\ \text{kHz} to 5 MHz5\ \text{MHz}) and watch the amplitude at the resistor's far end fall to 70 %70\,\% near 1.6 MHz1.6\ \text{MHz} with the 10x probe and near 160 kHz160\ \text{kHz} with 1x (within about ±30 %\pm 30\,\%, depending on your cable). Finally, view a fast generator edge with the long ground clip and with a spring tip: the ringing should mostly disappear.

Hazards and safe setup. Generator and calibrator levels only. The ground clip goes to circuit ground; no mains-referenced circuits.

Pitfalls

  • Using a probe compensated on another scope or channel.
  • Setting the scope channel to 1x with a 10x probe (readings ten times too small) or the reverse.
  • Measuring fast edges with a long ground lead and blaming the circuit for the ringing.
  • Clipping a grounded probe to a non-ground node; use a differential probe or two channels in math mode (A − B) for small signals only.

Practice

  1. A scope's input is 1 MΩ∥20 pF1\ \text{M}\Omega \parallel 20\ \text{pF} and the cable adds 60 pF60\ \text{pF}. What C1C_1 compensates a 9 MΩ9\ \text{M}\Omega tip?
  2. A probe with 8 pF8\ \text{pF} tip capacitance and a 100 nH100\ \text{nH} ground lead: at what frequency does it ring?
Answer
  1. C2=80 pFC_2 = 80\ \text{pF}, C1=R2C2/R1=80/9=8.9 pFC_1 = R_2 C_2/R_1 = 80/9 = 8.9\ \text{pF}.
  2. f=1/(2π100 nH×8 pF)=178 MHzf = 1/(2\pi\sqrt{100\ \text{nH} \times 8\ \text{pF}}) = 178\ \text{MHz}.

Test yourself

A 10x probe is correctly compensated when

  1. R1=R2R_1 = R_2
  2. R1C1=R2C2R_1 C_1 = R_2 C_2
  3. C1=C2C_1 = C_2
  4. the trimmer is at mid-travel

At 100 MHz the loading of a 10x passive probe is dominated by

  1. its 10 M ohm resistance
  2. its tip capacitance, about 160 Ω160\ \Omega of reactance
  3. the ground lead resistance
  4. the scope's ADC

A square wave shows ringing at about 100 MHz only when the long ground clip is used. The cause is

  1. the circuit oscillates
  2. the ground lead inductance resonating with the tip capacitance
  3. the probe is over-compensated
  4. aliasing

To view the voltage across a high-side shunt you should use

  1. a 10x probe with its ground clip on the shunt's low end
  2. a differential probe
  3. a 1x probe
  4. a current probe on the scope's ground