Measurements Exercises

These exercises cover the whole course in its order. Work each one by hand first, with units on every number, then check it with a short Python calculation or a simulation. Each answer links back to the lesson to revisit.

Fundamentals and uncertainty

1. Agreement. Two labs report 5.012±0.010 V5.012 \pm 0.010\ \text{V} and 4.996±0.012 V4.996 \pm 0.012\ \text{V} (k=2k = 2). Do they agree?

Answer

En=0.016/0.0102+0.0122=0.016/0.0156=1.02E_n = 0.016/\sqrt{0.010^2 + 0.012^2} = 0.016/0.0156 = 1.02: just above 1, so they do not agree at the stated uncertainties; look for an unrecognised bias or an underestimated uncertainty. Review: Introduction.

2. Static characteristics. A 00 to 100 kPa100\ \text{kPa} sensor with a 0.50.5 to 4.5 V4.5\ \text{V} output has a total error band of ±0.5 %\pm 0.5\,\% FS. What is the error limit in kPa, and what is it as a percentage of a reading at 10 kPa10\ \text{kPa}?

Answer

0.005×100 kPa=0.5 kPa0.005 \times 100\ \text{kPa} = 0.5\ \text{kPa}, which is 5 %5\,\% of a 10 kPa10\ \text{kPa} reading. Review: Static Characteristics.

3. Loading. A 47 kΩ47\ \text{k}\Omega / 47 kΩ47\ \text{k}\Omega divider from 12 V12\ \text{V} is read by a 10 MΩ10\ \text{M}\Omega DMM and by a 1 MΩ1\ \text{M}\Omega scope input. What does each read?

Answer

VTh=6 VV_\text{Th} = 6\ \text{V}, RTh=23.5 kΩR_\text{Th} = 23.5\ \text{k}\Omega. DMM: 6×10/10.0235=5.986 V6 \times 10/10.0235 = 5.986\ \text{V} (−0.23 %-0.23\,\%); scope: 6×1/1.0235=5.862 V6 \times 1/1.0235 = 5.862\ \text{V} (−2.3 %-2.3\,\%). Review: Errors and Loading.

4. Repeated readings. Six readings have s=0.80 mVs = 0.80\ \text{mV}. Give the standard uncertainty of their mean and its 95 %95\,\% half-width.

Answer

s/6=0.33 mVs/\sqrt{6} = 0.33\ \text{mV}; t0.975,5=2.571t_{0.975,5} = 2.571, so ±0.84 mV\pm 0.84\ \text{mV}. Review: Statistics of Repeated Readings.

5. Propagation. P=V2/RP = V^2/R with u(V)/V=0.10 %u(V)/V = 0.10\,\% and u(R)/R=0.20 %u(R)/R = 0.20\,\%, independent. Find u(P)/Pu(P)/P. What changes if VV and RR were measured with the same meter's gain error dominating both? (Think about which way it enters.)

Answer

(2×0.10)2+0.202=0.28 %\sqrt{(2 \times 0.10)^2 + 0.20^2} = 0.28\,\%. With a common gain error gg, V2/RV^2/R scales as (1+g)2/(1+g)=1+g(1 + g)^2/(1 + g) = 1 + g: only one power survives, so correlation changes the result; write the model with the shared error explicitly. Review: GUM Budgets, Propagation and Monte Carlo.

Meters and current

6. Shunt. A 50 mV50\ \text{mV} shunt rated 20 A20\ \text{A}: resistance and dissipation at full current?

Answer

2.5 mΩ2.5\ \text{m}\Omega; 202×0.0025=1.0 W20^2 \times 0.0025 = 1.0\ \text{W}. Review: Current Measurement.

7. True RMS. A triangle wave of 2 V2\ \text{V} peak is read by an average-responding meter and a true RMS meter. What do they show?

Answer

Rectified mean 1.000 V1.000\ \text{V}, average-responding reading 1.111 V1.111\ \text{V}; true RMS 2/3=1.155 V2/\sqrt{3} = 1.155\ \text{V}. The cheap meter is 3.8 %3.8\,\% low. Review: True RMS.

8. Dual slope. T1=40 msT_1 = 40\ \text{ms}, Vref=2.000 VV_\text{ref} = 2.000\ \text{V}, run-down t2=25.0 mst_2 = 25.0\ \text{ms}. What is VinV_\text{in}, and does T1T_1 reject 50 Hz50\ \text{Hz}?

Answer

Vin=2.000×25/40=1.250 VV_\text{in} = 2.000 \times 25/40 = 1.250\ \text{V}; 40 ms40\ \text{ms} is two cycles of 50 Hz50\ \text{Hz}, so yes (a null). Review: Inside the DMM.

9. Analog meter. A 100 μA100\ \mu\text{A}, 1 kΩ1\ \text{k}\Omega movement must read 10 V10\ \text{V} full scale. Find the multiplier and the sensitivity.

Answer

Rs=10/10−4−1000=99 kΩR_s = 10/10^{-4} - 1000 = 99\ \text{k}\Omega; 10 kΩ/V10\ \text{k}\Omega/\text{V}. Review: Analog Meters.

Resistance and impedance

10. Bridge. A balanced Wheatstone bridge has R1=100 ΩR_1 = 100\ \Omega, R2=1000 ΩR_2 = 1000\ \Omega, R3=120.0 ΩR_3 = 120.0\ \Omega. Find RxR_x.

Answer

Rx=R3R2/R1=1200 ΩR_x = R_3 R_2/R_1 = 1200\ \Omega. Review: DC Bridges.

11. Four-wire. Why does a 0.5 Ω0.5\ \Omega lead resistance not matter in a four-wire measurement, and what limits it if the leads get very long?

Answer

The sense leads carry almost no current, so they drop no voltage; very long force leads may exceed the current source's compliance voltage. Review: Four-Wire Resistance.

12. LCR. An LCR meter reads Cs=2.00 μFC_s = 2.00\ \mu\text{F} and D=0.30D = 0.30 at 10 kHz10\ \text{kHz}. What would the parallel model report?

Answer

Cp=Cs/(1+D2)=2.00/1.09=1.83 μFC_p = C_s/(1 + D^2) = 2.00/1.09 = 1.83\ \mu\text{F}. Review: LCR Meters; bridges: AC Bridges.

Oscilloscope and signal instruments

13. Rise time. A 2.0 ns2.0\ \text{ns} edge on a 200 MHz200\ \text{MHz} Gaussian scope: what rise time is displayed?

Answer

tr,scope=0.35/200 MHz=1.75 nst_{r,\text{scope}} = 0.35/200\ \text{MHz} = 1.75\ \text{ns}; 2.02+1.752=2.66 ns\sqrt{2.0^2 + 1.75^2} = 2.66\ \text{ns}. Review: Scope Acquisition.

14. Probe loading. What is the reactance of a 10 pF10\ \text{pF} probe tip at 5 MHz5\ \text{MHz}?

Answer

1/(2π×5×106×10−11)=3.18 kΩ1/(2\pi \times 5 \times 10^6 \times 10^{-11}) = 3.18\ \text{k}\Omega: a heavy load on a 1 kΩ1\ \text{k}\Omega node. Review: Probes.

15. FFT. A scope records 2 ms2\ \text{ms} at 500 MSa/s500\ \text{MSa/s}. What is the FFT's bin width and span?

Answer

Δf=1/2 ms=500 Hz\Delta f = 1/2\ \text{ms} = 500\ \text{Hz}; span 00 to 250 MHz250\ \text{MHz} (10610^6 points). Review: Triggering and FFT.

16. Counter. A reciprocal counter with 1 ns1\ \text{ns} resolution and a 0.1 s0.1\ \text{s} gate: what relative resolution? Why might its reading still be wrong by 1 ppm1\ \text{ppm}?

Answer

10−9/0.1=10−810^{-9}/0.1 = 10^{-8}; the internal timebase (offset, ageing, temperature) is typically ppm-level unless an external reference is used. Review: Generators and Counters.

17. dBm. A 1.0 V1.0\ \text{V} peak sine into 50 Ω50\ \Omega is how many dBm?

Answer

Vrms=0.707 VV_\text{rms} = 0.707\ \text{V}, P=0.5/50=10 mWP = 0.5/50 = 10\ \text{mW}, +10 dBm+10\ \text{dBm}. Review: Spectrum and Network Analysers.

Power, sensors, acquisition

18. THD (calculation only). A mains voltage has 5th, 7th and 11th harmonics of 5 %5\,\%, 3 %3\,\% and 2 %2\,\% of the fundamental. Find the THD.

Answer

25+9+4 %=6.2 %\sqrt{25 + 9 + 4}\,\% = 6.2\,\%, below the 8 %8\,\% THD limit that IEEE 519-2022 sets at the point of common coupling for ≤1 kV\le 1\ \text{kV}, and the 55th harmonic is just at its 5 %5\,\% individual limit. Review: Power Measurement.

19. Thermocouple. A type K reads 3.096 mV3.096\ \text{mV} with its terminals at 25 ∘C25\,^\circ\text{C} (E=1.000 mVE = 1.000\ \text{mV}). What is the hot-junction temperature?

Answer

3.096+1.000=4.096 mV3.096 + 1.000 = 4.096\ \text{mV}, which is 100 ∘C100\,^\circ\text{C} in the type K table. Review: Temperature Sensors.

20. Quarter bridge. A 120 Ω120\ \Omega gauge with GF 2.02.0 in a quarter bridge at 3 V3\ \text{V} sees 1000 με1000\ \mu\varepsilon. Estimate the output.

Answer

Vex GF ε/4=3×2×10−3/4=1.5 mVV_\text{ex}\,\text{GF}\,\varepsilon/4 = 3 \times 2 \times 10^{-3}/4 = 1.5\ \text{mV}. Review: Strain and Pressure.

21. Encoder. How many lines does an incremental encoder need for 0.1∘0.1^\circ with x4 decoding?

Answer

36003600 counts per revolution, 900900 lines. Review: Position, Light and Hall.

22. CMRR. An in-amp with 100 dB100\ \text{dB} CMRR sees 10 V10\ \text{V} of common mode. What error referred to the input?

Answer

10 V/105=100 μV10\ \text{V}/10^5 = 100\ \mu\text{V}. Review: Signal Conditioning.

23. ENOB. An ADC measures 70.0 dB70.0\ \text{dB} SINAD. What is its ENOB?

Answer

(70.0−1.76)/6.02=11.3(70.0 - 1.76)/6.02 = 11.3 bits. Review: Data Acquisition.

24. Ground loop. Why is lifting an instrument's protective earth never an acceptable cure for hum, and what are two acceptable ones?

Answer

The earth makes a fault safe; removing it can leave the chassis live. Use a differential or isolated measurement, or a single-point grounding plan (or a battery-powered instrument). Review: Grounding and Shielding.

Calibration, automation and reports

25. TUR. A tolerance of ±0.02 %\pm 0.02\ \% is verified with a calibrator of U=0.004 %U = 0.004\,\% (k=2k = 2). TUR?

Answer

0.02/0.004=5:10.02/0.004 = 5{:}1, above the usual 4:14{:}1. Review: Calibration.

26. SCPI. Put these in a safe order for a supply: OUTP ON, *RST, VOLT 5, CURR 0.05, SYST:ERR?.

Answer

*RST, CURR 0.05, VOLT 5, SYST:ERR?, OUTP ON (limits set and errors checked before the output goes live). Review: Automated Test.

27. Decision. Tolerance 10.00±0.10 mA10.00 \pm 0.10\ \text{mA}, reading 10.08 mA10.08\ \text{mA}, U=0.04 mAU = 0.04\ \text{mA} (k=2k = 2). Simple acceptance and guard band w=Uw = U?

Answer

Simple acceptance: pass. Guard band: acceptance zone 9.949.94 to 10.06 mA10.06\ \text{mA}, rejection beyond 10.14 mA10.14\ \text{mA}: inconclusive. pc=Φ((10.10−10.08)/0.02)=Φ(1)=84 %p_c = \Phi((10.10 - 10.08)/0.02) = \Phi(1) = 84\,\%. Review: Measurement Reports.

Check several answers in Python

python · try it yourself
import math
from scipy.stats import norm

print(f"1.  E_n = {0.016 / math.hypot(0.010, 0.012):.2f}")
print(f"3.  DMM {6 * 10e6 / (10e6 + 23.5e3):.3f} V, scope {6 * 1e6 / (1e6 + 23.5e3):.3f} V")
print(f"13. displayed rise time {math.hypot(2.0, 0.35 / 0.2):.2f} ns")
print(f"17. {10 * math.log10((0.5 / 50) / 1e-3):+.1f} dBm")
print(f"18. THD {100 * math.sqrt(0.05**2 + 0.03**2 + 0.02**2):.1f} %")
print(f"23. ENOB {(70.0 - 1.76) / 6.02:.1f} bits")
print(f"27. P(conform) = {100 * (norm.cdf(10.10, 10.08, 0.02) - norm.cdf(9.90, 10.08, 0.02)):.0f} %")
1.  E_n = 1.02
3.  DMM 5.986 V, scope 5.862 V
13. displayed rise time 2.66 ns
17. +10.0 dBm
18. THD 6.2 %
23. ENOB 11.3 bits
27. P(conform) = 84 %

Test yourself

Which error does averaging many readings NOT reduce?

  1. thermal noise
  2. quantisation noise of a busy signal
  3. a calibration offset
  4. random interference

Which measurement needs four wires most?

  1. 10 MΩ10\ \text{M}\Omega insulation
  2. a 10 mΩ10\ \text{m}\Omega shunt
  3. a 10 kΩ10\ \text{k}\Omega resistor
  4. a 5 V supply voltage

A ratiometric bridge measurement cancels

  1. the gauge factor
  2. excitation drift
  3. temperature effects on the gauge
  4. thermal EMFs

Traceability of a bench reading to the SI requires

  1. an expensive instrument
  2. an unbroken chain of calibrations with stated uncertainties
  3. a SCPI interface
  4. an ISO 9001 certificate