Home › Electrical & Electronic Measurements › DC Bridges: Wheatstone and Kelvin Double Bridge

DC Bridges: Wheatstone and Kelvin Double Bridge

Before precision DMMs, the most accurate way to measure a resistance was to compare it with a known one and adjust until a detector read zero. The method is still at the heart of measurement: strain-gauge and pressure sensors are bridges, resistance thermometers are read in bridges, and national laboratories compare resistance standards with bridge-like ratio instruments. The reason is that a null measurement does not depend on the detector's calibration or on the supply voltage, only on the ratio of the arms. The circuit analysis of bridges is in Circuit Theory; this lesson is about using them to measure.

The Wheatstone bridge as a measuring instrument

+ - Vs 5 V a R1 1 kOhm R2 1 kOhm R3 decade Rx unknown b c D Vd = vc - vb balance: Rx = R3 R2 / R1

Two dividers fed from the same source VsV_s: R1R_1 over R2R_2, and R3R_3 over the unknown RxR_x. The detector between the midpoints reads

Vd=vc−vb=Vs(RxR3+Rx−R2R1+R2)V_d = v_c - v_b = V_s\left(\frac{R_x}{R_3 + R_x} - \frac{R_2}{R_1 + R_2}\right)

It is zero, the bridge is balanced, when

R1R2=R3Rx⟹Rx=R3 R2R1\frac{R_1}{R_2} = \frac{R_3}{R_x} \quad\Longrightarrow\quad R_x = R_3\,\frac{R_2}{R_1}
  • VsV_s does not appear in the balance condition: supply drift and noise cancel (a perfectly correlated factor, Propagation).
  • The detector only has to recognise zero; its gain error is irrelevant. Its sensitivity decides how close to balance you can get.
  • The result's uncertainty comes from R3R_3 (usually a calibrated decade box) and the ratio R2/R1R_2/R_1 (which can be calibrated by swapping the arms).

Sensitivity near balance. With equal arms RR and Rx=R(1+x)R_x = R(1 + x) for a small fraction xx:

Vd≈Vs4 xV_d \approx \frac{V_s}{4}\,x

A 5 V5\ \text{V} bridge of 4.7 kΩ4.7\ \text{k}\Omega arms gives 5/4×(1/4700)=0.27 mV5/4 \times (1/4700) = 0.27\ \text{mV} per ohm of imbalance: a microvolt detector resolves a few milliohms. Raising VsV_s raises the sensitivity until self-heating of the arms becomes the limit.

Worked example: balancing an unknown resistor

Given: Vs=5.000 VV_s = 5.000\ \text{V}, R1=R2=1.000 kΩR_1 = R_2 = 1.000\ \text{k}\Omega (ratio 1.00001.0000, calibrated), R3R_3 a decade box set to 4700 Ω4700\ \Omega, the unknown RxR_x near 4.7 kΩ4.7\ \text{k}\Omega. The detector reads Vd=+8.22 mVV_d = +8.22\ \text{mV}.

Find: RxR_x from the imbalance, and the decade setting for balance.

Solve: vb=2.500 Vv_b = 2.500\ \text{V}, so vc=2.50822 Vv_c = 2.50822\ \text{V} and

RxR3+Rx=2.508225.000=0.501644Rx=R3 0.5016441−0.501644=4700 Ω×1.006597=4731.0 Ω\begin{aligned} \frac{R_x}{R_3 + R_x} &= \frac{2.50822}{5.000} = 0.501644 \\ R_x &= R_3\,\frac{0.501644}{1 - 0.501644} = 4700\ \Omega \times 1.006597 = 4731.0\ \Omega \end{aligned}

At balance the decade reads 4731 Ω4731\ \Omega.

Check: the small-signal estimate x≈4Vd/Vs=4×8.22/5000=0.66 %x \approx 4V_d/V_s = 4 \times 8.22/5000 = 0.66\,\% gives 4731 Ω4731\ \Omega too. The imbalance method needs the detector and VsV_s calibrated; the null method needs neither, which is why you finish by balancing.

Simulation. R3 is the decade box and Rx the unknown (4731 Ω4731\ \Omega). E1 buffers the detector voltage vc−vbv_c - v_b onto node d. At R3=4.7 kΩR_3 = 4.7\ \text{k}\Omega node d reads 8.22 mV8.22\ \text{mV}; slide R3R_3 to 4731 Ω4731\ \Omega and it reads zero. Notice how the output changes sign on either side of balance, which tells the operator which way to turn the dial.

node voltages (to ground 0)a5.000 Vb2.500 Vc2.508 Vd8.218 mVnetlistV1 a 0 DC 5R1 a b 1kR2 b 0 1kR3 a c 4.7kRx c 0 4731E1 d 0 c b 1Rd d 0 1megelement currents (node 1 to node 2 through it)V1a>0-3.03 mAR1a>b2.50 mAR2b>02.50 mAR3a>c530 uARxc>0530 uAE1d>0-8.22 nARdd>08.22 nA

Low resistance: the Kelvin double bridge

For resistances below about 1 Ω1\ \Omega the Wheatstone bridge fails for the same reason a 2-wire ohmmeter does: the leads, contacts and the link joining the unknown to the standard are in series with them (Four-Wire Resistance). The Kelvin double bridge (Thomson bridge) passes a large current through the unknown RxR_x, the link rr and a low-value standard RR in series, and uses two pairs of ratio arms: outer arms PP, QQ to the outer potential terminals, inner arms pp, qq to the inner ones on each side of the link. At balance

Rx=R PQ+q rp+q+r(PQ−pq)R_x = R\,\frac{P}{Q} + \frac{q\,r}{p + q + r}\left(\frac{P}{Q} - \frac{p}{q}\right)

If the two ratios are made equal (p/q=P/Qp/q = P/Q, usually by ganging them), the second term vanishes and the link resistance drops out entirely. A small mismatch leaves a small correction proportional to rr.

python · try it yourself
import numpy as np
from scipy.optimize import brentq

def solve(R, P, Q, p, q, r, Rx, I=10.0):
    """Kelvin double bridge, nodal analysis. Nodes: a (Rx outer, current in), b (Rx inner), c (R inner),
    d (R outer, current out, ground), g (P-Q junction), h (p-q junction). Returns detector voltage v_g - v_h."""
    nodes = ["a", "b", "c", "g", "h"]          # d is ground
    idx = {n: i for i, n in enumerate(nodes)}
    G = np.zeros((5, 5))
    J = np.zeros(5)
    def add(n1, n2, res):
        g = 1 / res
        for n in (n1, n2):
            if n != "d":
                G[idx[n], idx[n]] += g
        if n1 != "d" and n2 != "d":
            G[idx[n1], idx[n2]] -= g
            G[idx[n2], idx[n1]] -= g
    add("a", "b", Rx); add("b", "c", r); add("c", "d", R)
    add("a", "g", P); add("g", "d", Q); add("b", "h", p); add("h", "c", q)
    J[idx["a"]] = I                              # test current into a, out of d
    v = np.linalg.solve(G, J)
    return v[idx["g"]] - v[idx["h"]]

R, r = 1e-3, 0.5e-3                              # standard 1 mohm, link 0.5 mohm
P, Q = 40.0, 100.0
print("p/q matched:")
for Rx in (0.39e-3, 0.40e-3, 0.41e-3):
    print(f"  Rx = {1e3 * Rx:.2f} mohm: detector {round(1e6 * solve(R, P, Q, 40.0, 100.0, r, Rx), 3) + 0.0:+8.3f} uV")
# Balance point with a 1 % mismatch of the inner ratio p/q
Rx_bal = brentq(lambda x: solve(R, P, Q, 40.4, 100.0, r, x), 0.3e-3, 0.5e-3)
corr = 100.0 * r / (40.4 + 100.0 + r) * (P / Q - 40.4 / 100.0)
print(f"p 1 % high: balance at Rx = {1e3 * Rx_bal:.5f} mohm (R P/Q = 0.40000), "
      f"error {100 * (Rx_bal / 0.4e-3 - 1):+.3f} %, formula correction {1e3 * corr:+.5f} mohm")
print(f"plain Wheatstone (link in series with Rx): would read {1e3 * (0.4e-3 + r):.2f} mohm")
p/q matched:
  Rx = 0.39 mohm: detector  -71.428 uV
  Rx = 0.40 mohm: detector   +0.000 uV
  Rx = 0.41 mohm: detector  +71.428 uV
p 1 % high: balance at Rx = 0.39858 mohm (R P/Q = 0.40000), error -0.356 %, formula correction -0.00142 mohm
plain Wheatstone (link in series with Rx): would read 0.90 mohm

With matched ratios the detector is exactly zero at Rx=0.400 mΩR_x = 0.400\ \text{m}\Omega although the link (0.5 mΩ0.5\ \text{m}\Omega) is larger than the unknown. A 1 %1\,\% mismatch of the inner arms moves the balance by 0.36 %0.36\,\%, exactly the formula's correction term. A Wheatstone bridge would have included the link and read 0.90 mΩ0.90\ \text{m}\Omega, more than twice the true value. Modern micro-ohmmeters do the same job with a four-wire current source and a nanovoltmeter, using current reversal for thermal EMFs.

Guided lab: balance a bridge with a decade box

Hand calculation. With R1=R2=1 kΩR_1 = R_2 = 1\ \text{k}\Omega (0.1 %0.1\,\%), a 4.7 kΩ4.7\ \text{k}\Omega, 1 %1\,\% unknown should balance somewhere between 46534653 and 4747 Ω4747\ \Omega; the result's uncertainty is the decade box's accuracy plus the R2/R1R_2/R_1 ratio's (±0.2 %\pm 0.2\,\% worst case, less after swapping the arms and averaging).

Simulation. The netlist lab above, or edit Rx to your resistor's nominal value.

What you would measure. Build the bridge on a breadboard from a 5 V5\ \text{V} supply with a decade resistance box as R3R_3 and a DMM on its 100 mV100\ \text{mV} range as the detector (floating across b and c). Expect a detector slope of about 0.27 mV0.27\ \text{mV} per ohm near balance and a balance setting within the unknown's tolerance. Then swap R1R_1 and R2R_2, rebalance, and average the two settings: the ratio error of the arms cancels to first order.

Hazards and safe setup. 5 V5\ \text{V}, current limit 20 mA20\ \text{mA}; each arm dissipates at most about 6 mW6\ \text{mW}. The detector DMM must be floating (battery-powered or a bench DMM with isolated inputs), since neither midpoint is at ground.

Pitfalls

  • Using a Wheatstone bridge for milliohms: the link and contacts dominate.
  • Grounding one side of the detector with a scope probe, which shorts an arm through earth.
  • Raising the bridge voltage for sensitivity and heating the arms (their resistance changes).
  • Forgetting thermal EMFs in low-resistance bridges: reverse the current and average.

Practice

  1. A balanced bridge has R1=100 ΩR_1 = 100\ \Omega, R2=1000 ΩR_2 = 1000\ \Omega, R3=562.4 ΩR_3 = 562.4\ \Omega. What is RxR_x?
  2. A 10 V10\ \text{V} bridge of 350 Ω350\ \Omega arms is unbalanced by 0.35 Ω0.35\ \Omega in one arm. Estimate VdV_d.
Answer
  1. Rx=R3R2/R1=562.4×10=5624 ΩR_x = R_3 R_2/R_1 = 562.4 \times 10 = 5624\ \Omega.
  2. x=0.001x = 0.001, Vd≈10/4×0.001=2.5 mVV_d \approx 10/4 \times 0.001 = 2.5\ \text{mV}.

Test yourself

The balance condition of a Wheatstone bridge depends on

  1. the supply voltage
  2. the detector's gain
  3. only the ratios of the arm resistances
  4. the detector's input resistance

Near balance, a bridge of equal arms with supply VsV_s gives an output of about

  1. VsxV_s x
  2. Vsx/2V_s x/2
  3. Vsx/4V_s x/4
  4. Vs/xV_s/x

The Kelvin double bridge removes the effect of

  1. the supply voltage
  2. the link and contact resistance between unknown and standard
  3. the detector's offset
  4. the temperature coefficient of the unknown

Why finish a bridge measurement by nulling rather than reading the imbalance?

  1. it is faster
  2. the null does not depend on the detector's calibration or the supply value
  3. the imbalance is always zero
  4. the detector saturates otherwise