DC Bridges: Wheatstone and Kelvin Double Bridge
Before precision DMMs, the most accurate way to measure a resistance was to compare it with a known one and adjust until a detector read zero. The method is still at the heart of measurement: strain-gauge and pressure sensors are bridges, resistance thermometers are read in bridges, and national laboratories compare resistance standards with bridge-like ratio instruments. The reason is that a null measurement does not depend on the detector's calibration or on the supply voltage, only on the ratio of the arms. The circuit analysis of bridges is in Circuit Theory; this lesson is about using them to measure.
The Wheatstone bridge as a measuring instrument
Two dividers fed from the same source : over , and over the unknown . The detector between the midpoints reads
It is zero, the bridge is balanced, when
- does not appear in the balance condition: supply drift and noise cancel (a perfectly correlated factor, Propagation).
- The detector only has to recognise zero; its gain error is irrelevant. Its sensitivity decides how close to balance you can get.
- The result's uncertainty comes from (usually a calibrated decade box) and the ratio (which can be calibrated by swapping the arms).
Sensitivity near balance. With equal arms and for a small fraction :
A bridge of arms gives per ohm of imbalance: a microvolt detector resolves a few milliohms. Raising raises the sensitivity until self-heating of the arms becomes the limit.
Worked example: balancing an unknown resistor
Given: , (ratio , calibrated), a decade box set to , the unknown near . The detector reads .
Find: from the imbalance, and the decade setting for balance.
Solve: , so and
At balance the decade reads .
Check: the small-signal estimate gives too. The imbalance method needs the detector and calibrated; the null method needs neither, which is why you finish by balancing.
Simulation. R3 is the decade box and Rx the unknown (). E1 buffers the detector voltage onto node d. At node d reads ; slide to and it reads zero. Notice how the output changes sign on either side of balance, which tells the operator which way to turn the dial.
Low resistance: the Kelvin double bridge
For resistances below about the Wheatstone bridge fails for the same reason a 2-wire ohmmeter does: the leads, contacts and the link joining the unknown to the standard are in series with them (Four-Wire Resistance). The Kelvin double bridge (Thomson bridge) passes a large current through the unknown , the link and a low-value standard in series, and uses two pairs of ratio arms: outer arms , to the outer potential terminals, inner arms , to the inner ones on each side of the link. At balance
If the two ratios are made equal (, usually by ganging them), the second term vanishes and the link resistance drops out entirely. A small mismatch leaves a small correction proportional to .
import numpy as np
from scipy.optimize import brentq
def solve(R, P, Q, p, q, r, Rx, I=10.0):
"""Kelvin double bridge, nodal analysis. Nodes: a (Rx outer, current in), b (Rx inner), c (R inner),
d (R outer, current out, ground), g (P-Q junction), h (p-q junction). Returns detector voltage v_g - v_h."""
nodes = ["a", "b", "c", "g", "h"] # d is ground
idx = {n: i for i, n in enumerate(nodes)}
G = np.zeros((5, 5))
J = np.zeros(5)
def add(n1, n2, res):
g = 1 / res
for n in (n1, n2):
if n != "d":
G[idx[n], idx[n]] += g
if n1 != "d" and n2 != "d":
G[idx[n1], idx[n2]] -= g
G[idx[n2], idx[n1]] -= g
add("a", "b", Rx); add("b", "c", r); add("c", "d", R)
add("a", "g", P); add("g", "d", Q); add("b", "h", p); add("h", "c", q)
J[idx["a"]] = I # test current into a, out of d
v = np.linalg.solve(G, J)
return v[idx["g"]] - v[idx["h"]]
R, r = 1e-3, 0.5e-3 # standard 1 mohm, link 0.5 mohm
P, Q = 40.0, 100.0
print("p/q matched:")
for Rx in (0.39e-3, 0.40e-3, 0.41e-3):
print(f" Rx = {1e3 * Rx:.2f} mohm: detector {round(1e6 * solve(R, P, Q, 40.0, 100.0, r, Rx), 3) + 0.0:+8.3f} uV")
# Balance point with a 1 % mismatch of the inner ratio p/q
Rx_bal = brentq(lambda x: solve(R, P, Q, 40.4, 100.0, r, x), 0.3e-3, 0.5e-3)
corr = 100.0 * r / (40.4 + 100.0 + r) * (P / Q - 40.4 / 100.0)
print(f"p 1 % high: balance at Rx = {1e3 * Rx_bal:.5f} mohm (R P/Q = 0.40000), "
f"error {100 * (Rx_bal / 0.4e-3 - 1):+.3f} %, formula correction {1e3 * corr:+.5f} mohm")
print(f"plain Wheatstone (link in series with Rx): would read {1e3 * (0.4e-3 + r):.2f} mohm")p/q matched:
Rx = 0.39 mohm: detector -71.428 uV
Rx = 0.40 mohm: detector +0.000 uV
Rx = 0.41 mohm: detector +71.428 uV
p 1 % high: balance at Rx = 0.39858 mohm (R P/Q = 0.40000), error -0.356 %, formula correction -0.00142 mohm
plain Wheatstone (link in series with Rx): would read 0.90 mohm
With matched ratios the detector is exactly zero at although the link () is larger than the unknown. A mismatch of the inner arms moves the balance by , exactly the formula's correction term. A Wheatstone bridge would have included the link and read , more than twice the true value. Modern micro-ohmmeters do the same job with a four-wire current source and a nanovoltmeter, using current reversal for thermal EMFs.
Guided lab: balance a bridge with a decade box
Hand calculation. With (), a , unknown should balance somewhere between and ; the result's uncertainty is the decade box's accuracy plus the ratio's ( worst case, less after swapping the arms and averaging).
Simulation. The netlist lab above, or edit Rx to your resistor's nominal value.
What you would measure. Build the bridge on a breadboard from a supply with a decade resistance box as and a DMM on its range as the detector (floating across b and c). Expect a detector slope of about per ohm near balance and a balance setting within the unknown's tolerance. Then swap and , rebalance, and average the two settings: the ratio error of the arms cancels to first order.
Hazards and safe setup. , current limit ; each arm dissipates at most about . The detector DMM must be floating (battery-powered or a bench DMM with isolated inputs), since neither midpoint is at ground.
Pitfalls
- Using a Wheatstone bridge for milliohms: the link and contacts dominate.
- Grounding one side of the detector with a scope probe, which shorts an arm through earth.
- Raising the bridge voltage for sensitivity and heating the arms (their resistance changes).
- Forgetting thermal EMFs in low-resistance bridges: reverse the current and average.
Practice
- A balanced bridge has , , . What is ?
- A bridge of arms is unbalanced by in one arm. Estimate .
Answer
- .
- , .
Test yourself
The balance condition of a Wheatstone bridge depends on
- the supply voltage
- the detector's gain
- only the ratios of the arm resistances
- the detector's input resistance
Near balance, a bridge of equal arms with supply gives an output of about
The Kelvin double bridge removes the effect of
- the supply voltage
- the link and contact resistance between unknown and standard
- the detector's offset
- the temperature coefficient of the unknown
Why finish a bridge measurement by nulling rather than reading the imbalance?
- it is faster
- the null does not depend on the detector's calibration or the supply value
- the imbalance is always zero
- the detector saturates otherwise