TExcelTAD
Home › Electrical & Electronic Measurements › Two-Wire and Four-Wire Resistance, Thermal EMFs and Insulation Testing

Two-Wire and Four-Wire Resistance, Thermal EMFs and Insulation Testing

Measure a 1 Ω1\ \Omega current-sense resistor with a DMM's ohms function and you may read 1.3 Ω1.3\ \Omega. Touch the leads together first and you read 0.3 Ω0.3\ \Omega: the leads and contacts are almost a quarter of the reading. Subtracting that number (the "relative" or "null" function) helps, but contact resistance changes every time you press the probes. For low resistances the fix is to stop measuring the leads at all: four-wire (Kelvin) measurement. This lesson covers it, the thermal EMFs that limit it at the milliohm level, and, at the other end of the scale, the principle of insulation resistance testing.

Two wires: the leads are in series

An ohmmeter drives a known test current II through the unknown and measures the voltage. With two wires, the same leads carry the current and sense the voltage, so the meter sees

R2w=Rx+Rlead,1+Rlead,2+RcontactR_\text{2w} = R_x + R_\text{lead,1} + R_\text{lead,2} + R_\text{contact}

Typical test leads are 0.050.05 to 0.2 Ω0.2\ \Omega each including contacts. On a 10 kΩ10\ \text{k}\Omega resistor that is negligible; on 1 Ω1\ \Omega it is a 10 %10\,\% to 40 %40\,\% error.

Four wires: force and sense

I 10 mA Rf1 force lead Rf2 force lead Rx 1 Ohm Rs1 sense lead Rs2 sense lead V 10 MOhm I I sense ~ 0 I flows only in the force loop

Two force leads carry the test current; two separate sense leads connect to the voltmeter at the resistor's own terminals, inside the force connections. The voltmeter's input resistance is 10 MΩ10\ \text{M}\Omega or more, so the sense leads carry essentially no current and drop no voltage:

R4w=VsenseI=Rx(to within Rsense leads/Rin)R_\text{4w} = \frac{V_\text{sense}}{I} = R_x \quad \text{(to within } R_\text{sense leads}/R_\text{in}\text{)}

The lead resistance now matters only if it is so large that the current source runs out of compliance voltage. The same idea appears in four-terminal shunts, remote-sense power supplies, Kelvin-connected PCB footprints and the three- and four-wire RTD connections of Temperature Sensors.

Worked example: a 1 ohm resistor, 2-wire and 4-wire

Given: Rx=1.000 ΩR_x = 1.000\ \Omega; each of the four leads (with its contact) 0.15 Ω0.15\ \Omega; test current 10 mA10\ \text{mA}; voltmeter input 10 MΩ10\ \text{M}\Omega.

Find: the 2-wire and 4-wire readings.

Solve:

R2w=10 mA×(0.15+1.000+0.15) Ω10 mA=1.300 ΩR4w≈1.000 Ω×10 MΩ10 MΩ+0.3 Ω=0.99999997 Ω\begin{aligned} R_\text{2w} &= \frac{10\ \text{mA} \times (0.15 + 1.000 + 0.15)\ \Omega}{10\ \text{mA}} = 1.300\ \Omega \\ R_\text{4w} &\approx 1.000\ \Omega \times \frac{10\ \text{M}\Omega}{10\ \text{M}\Omega + 0.3\ \Omega} = 0.99999997\ \Omega \end{aligned}

Check: the 4-wire error is 3×10−83 \times 10^{-8} relative, far below any meter's specification. ngspice gives the same numbers, with the two measurement styles in one deck:

bash · try it yourself
cat > kelvin.cir <<'EOF'
* 1 ohm resistor measured by a 10 mA ohmmeter: 2-wire and 4-wire
I1 0 hf DC 10m
Rf1 hf a 0.15
Rx a b 1.0
Rf2 b 0 0.15
* sense leads to a 10 Mohm voltmeter between hs and ls
Rs1 a hs 0.15
Rs2 b ls 0.15
Rvm hs ls 10meg
.control
op
let r2w = v(hf) / 10m
let r4w = (v(hs) - v(ls)) / 10m
print r2w r4w
.endc
.end
EOF
ngspice -b kelvin.cir 2>&1 | grep -E '^r[24]w'
r2w = 1.300000e+00
r4w = 9.999999e-01

Simulation. The same circuit as a lab: Rf1 is the HI force lead; slide it from 0.10.1 to 5 Ω5\ \Omega. Node hf (what a 2-wire meter reads, divided by 10 mA10\ \text{mA}) moves with it; hs minus ls stays at 10.000 mV10.000\ \text{mV}, i.e. 1.000 Ω1.000\ \Omega.

node voltages (to ground 0)hf13.00 mVa11.50 mVb1.500 mVhs11.50 mVls1.500 mVnetlistI1 0 hf DC 10mRf1 hf a 0.15Rx a b 1Rf2 b 0 0.15Rs1 a hs 0.15Rs2 b ls 0.15Rvm hs ls 10megelement currents (node 1 to node 2 through it)I10>hf10.0 mARf1hf>a10.0 mARxa>b10.0 mARf2b>010.0 mARs1a>hs1.00 nARs2b>ls-1.00 nARvmhs>ls1.00 nA

Thermal EMFs and offset compensation

Wherever two different metals meet at a temperature difference, a thermocouple voltage appears (Temperature Sensors): copper to tin–lead solder gives a few μV/K\mu\text{V/K}, copper to oxidised copper far more. In a low-resistance measurement the signal is small (100 mA100\ \text{mA} through 10 mΩ10\ \text{m}\Omega is 1 mV1\ \text{mV}), so a few microvolts of thermal EMF is a large error. Two remedies:

  • Offset compensation: measure VV with the current on and off, and subtract.
  • Current reversal: measure with +I+I and −I-I; R=(V+−V−)/(2I)R = (V_+ - V_-)/(2I) cancels any constant EMF and doubles the signal.

Both assume the EMF is constant over the measurement; let connections reach thermal equilibrium after handling.

python · try it yourself
import numpy as np

# Four-wire measurement of a 10 mohm shunt with a 100 mA test current and a 7 uV thermal EMF.
Rx, I, E_th = 0.010, 0.100, 7e-6
rng = np.random.default_rng(2)
noise = lambda: 0.2e-6 * rng.standard_normal()      # 0.2 uV voltmeter noise per reading

v_plus = I * Rx + E_th + noise()                     # current forward
v_minus = -I * Rx + E_th + noise()                   # current reversed
v_off = E_th + noise()                               # current off
print(f"plain reading:             R = {1000 * v_plus / I:.4f} mohm")
print(f"offset-compensated (off):  R = {1000 * (v_plus - v_off) / I:.4f} mohm")
print(f"current reversal:          R = {1000 * (v_plus - v_minus) / (2 * I):.4f} mohm")

# Two-wire error for different leads (0.05 ohm per lead incl. contact)
for R in (1000.0, 10.0, 1.0, 0.1):
    print(f"2-wire on {R:7.1f} ohm with 2 x 0.05 ohm leads: error {100 * 0.1 / R:8.3f} %")
plain reading:             R = 10.0704 mohm
offset-compensated (off):  R = 10.0012 mohm
current reversal:          R = 10.0007 mohm
2-wire on  1000.0 ohm with 2 x 0.05 ohm leads: error    0.010 %
2-wire on    10.0 ohm with 2 x 0.05 ohm leads: error    1.000 %
2-wire on     1.0 ohm with 2 x 0.05 ohm leads: error   10.000 %
2-wire on     0.1 ohm with 2 x 0.05 ohm leads: error  100.000 %

A 7 μV7\ \mu\text{V} EMF adds 0.7 %0.7\,\% to the plain reading; both compensated methods remove it to within the voltmeter noise. The 2-wire table shows why every resistance below about 10 Ω10\ \Omega deserves four wires.

High resistance and insulation testing (concept)

At the other extreme, insulation resistance (between a cable's conductors, a motor winding and its frame, a transformer's windings) is in the megohm to teraohm range and is measured at a high test voltage, typically 250 V250\ \text{V}, 500 V500\ \text{V} or 1000 V1000\ \text{V} DC for low-voltage equipment, by an insulation tester. The tester reports R=V/IR = V/I from a nanoampere-level current; it applies a high voltage precisely because insulation can look perfect at 10 V10\ \text{V} and fail at its working voltage. Practice includes:

  • a guard terminal, which diverts surface leakage current around the current-measuring circuit, so only the leakage through the insulation is measured (the same three-terminal idea as in LCR Meters);
  • time readings such as the polarisation index (the ratio of the readings at 10 minutes and at 1 minute; a low ratio suggests moist or contaminated insulation);
  • discharging the item after the test, since cables and windings store charge.

This is a concept in this course, not a lab. Insulation testers produce hazardous voltages and are used by qualified people on de-energised, isolated equipment under the applicable standards (for installations, IEC 60364-6 sets the test voltages and minimum insulation resistances; the edition in force in 2026 is IEC 60364-6:2016, with a third edition in preparation, so check which edition applies). On the bench you can explore high resistance safely with a DMM's >10 GΩ>10\ \text{G}\Omega input or by timing the discharge of a capacitor through the unknown.

Guided lab: measure a 0.1 ohm resistor three ways

Hand calculation. A 0.1 Ω0.1\ \Omega, 1 %1\,\% resistor with 0.1 Ω0.1\ \Omega leads: 2-wire reads about 0.3 Ω0.3\ \Omega, 2-wire after "null" about 0.100.10 to 0.12 Ω0.12\ \Omega (contact variation), 4-wire 0.100±0.001 Ω0.100 \pm 0.001\ \Omega.

Simulation. Run the ngspice deck with Rx a b 0.1 and the lab with the lead slider.

What you would measure. Use a bench DMM's 4-wire function if it has one. If not, build it: a bench supply in constant-current mode at 100 mA100\ \text{mA} (current limit set first, voltage limit 1 V1\ \text{V}) through the resistor, a DMM on its mA range in series (force loop), and a second DMM on its 100 mV100\ \text{mV} range clipped to the resistor's leads right at its body (sense). Expect 10.0 mV10.0\ \text{mV} for 100 mA100\ \text{mA}, and a few microvolts of change when you reverse the supply leads (thermal EMF).

Hazards and safe setup. 100 mA100\ \text{mA} in 0.1 Ω0.1\ \Omega is 1 mW1\ \text{mW}; the supply's 1 V1\ \text{V} limit keeps any mistake harmless. No insulation tester on the bench.

Pitfalls

  • Clipping the sense leads outside the force leads (the contact resistance of the force clips then sits inside the measurement).
  • Ignoring compliance: a 4-wire source with long thin leads may not reach its test current.
  • Measuring milliohms without offset compensation.
  • Touching an item during or right after an insulation test.

Practice

  1. A 2-wire reading of a 2.2 Ω2.2\ \Omega resistor gives 2.47 Ω2.47\ \Omega; leads shorted give 0.21 Ω0.21\ \Omega. What is the best estimate, and what remains uncertain?
  2. With +I+I the voltage is 1.0123 mV1.0123\ \text{mV}, with −I-I it is −0.9877 mV-0.9877\ \text{mV}, I=100 mAI = 100\ \text{mA}. Find RR and the thermal EMF.
Answer
  1. 2.47−0.21=2.26 Ω2.47 - 0.21 = 2.26\ \Omega; the contact resistance changed between the two contacts, so the result is uncertain by several tens of milliohms. Use four wires.
  2. R=(1.0123+0.9877) mV/(0.2 A)=10.000 mΩR = (1.0123 + 0.9877)\ \text{mV}/(0.2\ \text{A}) = 10.000\ \text{m}\Omega; EMF =(1.0123−0.9877)/2=12.3 μV= (1.0123 - 0.9877)/2 = 12.3\ \mu\text{V}.

Test yourself

In a four-wire measurement the sense leads

  1. carry the test current
  2. carry almost no current, so their resistance drops no voltage
  3. must be shorter than the force leads
  4. are connected outside the force leads

Current reversal in a low-resistance measurement cancels

  1. lead resistance
  2. constant thermal EMFs
  3. self-heating
  4. the meter's gain error

Why are insulation tests made at hundreds of volts?

  1. to heat the insulation
  2. insulation weaknesses may only conduct at working-level voltages
  3. DMMs cannot measure megohms
  4. to discharge the cable

A 2-wire reading of 1 Ω1\ \Omega with 0.05 Ω0.05\ \Omega per lead has an error of about

  1. 0.1 %
  2. 1 %
  3. 10 %
  4. 50 %