Gross, Systematic, Random and Loading Errors
A technician measures the midpoint of a / divider fed from with an old analog multimeter and reads . A colleague's DMM reads . Neither meter is broken, and averaging a thousand readings from either would not bring them together. The difference is a loading error, a systematic error caused by the instrument itself. This lesson sorts measurement errors into four kinds, because each kind has a different cure, and works through loading in detail.
Four kinds of error
The error of a result is the result minus the true value. You never know it exactly (you would not need to measure), but you can know its causes.
| Kind | Examples | Detect it by | Cure |
|---|---|---|---|
| gross (blunders) | wrong range or function, leads in the A jack, misread scale, digits swapped in the notebook, wrong probe attenuation setting | sanity checks, a second method, plotting the data | procedure, checklists, automation; never "average out" |
| systematic | calibration offset and gain, loading, burden voltage, thermal EMFs, lead resistance, a temperature coefficient, parallax | comparison with a better instrument or a different method; changing the suspected cause | calibrate, correct, redesign the method; then assign an uncertainty to the correction |
| random | thermal noise, quantisation, interference that is not synchronous with the reading, operator reading jitter | scatter of repeated readings | averaging, filtering, longer integration, shielding |
| drift | warm-up, ageing, slow temperature change | readings that trend in time | warm-up time, recalibration interval, monitoring |
The distinction between systematic and random depends on the time scale: hum is random from reading to reading of a fast scope, but a fixed offset if every reading is synchronised to the mains. The GUM drops the words in favour of "Type A" and "Type B" evaluations (GUM Uncertainty Budgets), but the causes still matter when you decide what to fix.
Averaging removes only random error. The mean of readings has a random part that shrinks as ; the bias does not shrink at all.
import numpy as np
rng = np.random.default_rng(7)
true_V = 5.000
bias = 0.012 # meter gain/offset error, V (systematic)
noise = 0.020 # reading-to-reading standard deviation, V (random)
for N in (1, 10, 100, 1000, 10000):
means = [float(np.mean(true_V + bias + noise * rng.standard_normal(N))) for _ in range(200)]
spread = float(np.std(means))
print(f"N = {N:5d}: scatter of the mean {1000 * spread:6.2f} mV, "
f"mean error {1000 * (np.mean(means) - true_V):+6.2f} mV")N = 1: scatter of the mean 17.45 mV, mean error +9.36 mV
N = 10: scatter of the mean 6.86 mV, mean error +11.44 mV
N = 100: scatter of the mean 1.98 mV, mean error +11.95 mV
N = 1000: scatter of the mean 0.65 mV, mean error +12.03 mV
N = 10000: scatter of the mean 0.19 mV, mean error +11.99 mV
The scatter falls by per decade of ; the bias stays. Past the result is limited by the bias, and only calibration helps.
Loading: the instrument becomes part of the circuit
Any circuit seen from the two points you measure between is a Thevenin source behind (Thevenin and Norton). A voltmeter of input resistance forms a divider with it:
The relative error is about : to stay below the meter needs times the source resistance. Because the error is known, it can be corrected:
and the uncertainties of and then enter the budget instead of the whole error.
An ammeter is inserted in series, so its resistance (the shunt plus leads and fuse) adds to the loop. A source of behind a loop resistance gives without the meter and with it, an error of about . The voltage across the meter, , is its burden voltage (Current Measurement).
At AC the input is an impedance: a scope input in parallel with has at , so a source that is harmless to load at DC is heavily loaded at radio frequencies (Probes).
Worked example: the divider and two meters
Given: , ; a DMM with ; an analog meter rated on its range.
Find: both readings, their errors, and the corrected value from the analog reading.
Solve: , . The analog meter's input is .
Check: both readings are low, as loading always lowers a voltage reading on a resistive source. The analog meter's relative error is times the DMM's, not 50 (the ratio of the input resistances), because the approximation fails when the meter is only four times the source resistance. Units cancel in every ratio.
The same deck in ngspice, with the meter as a resistor Rm that the control script changes:
cat > load.cir <<'EOF'
* 100k/100k divider read by voltmeters of different input resistance
V1 in 0 DC 10
R1 in out 100k
R2 out 0 100k
Rm out 0 10meg
.control
op
echo "Rm = 10 Mohm (DMM)"
print v(out)
alter Rm = 1meg
op
echo "Rm = 1 Mohm (scope input)"
print v(out)
alter Rm = 200k
op
echo "Rm = 200 kohm (analog meter, 20 kohm/V on 10 V)"
print v(out)
.endc
.end
EOF
ngspice -b load.cir 2>&1 | grep -E '^(v\(|Rm)'Rm = 10 Mohm (DMM)
v(out) = 4.975124e+00
Rm = 1 Mohm (scope input)
v(out) = 4.761905e+00
Rm = 200 kohm (analog meter, 20 kohm/V on 10 V)
v(out) = 4.000000e+00
Guided lab: how high must the meter's resistance be?
Hand calculation. For the divider above, the error is : at , at , at .
Simulation. Rm is the meter. Slide it from to and read node out: at , at , at . A / divider () would err by only at and at : low source resistance is the real cure.
What you would measure. With resistors and a supply set with the DMM itself: a DMM reading of about (within from resistor tolerance), a scope reading near (if your scope has input and a 1x probe; a 10x probe gives and about ). Some DMMs switch to a input on their lowest DC ranges; on those the loading disappears, which is itself a useful check of the input-resistance setting.
Hazards and safe setup. Current-limited supply; the divider draws . No hazard beyond the usual: the scope's probe ground goes to circuit ground only.
Pitfalls
- Averaging to "improve accuracy" when the error is systematic.
- Forgetting that the input resistance changes with range and function (analog meters: per volt of range; some DMMs: or selectable).
- Correcting for loading but leaving the correction out of the uncertainty budget.
- Assuming a DC input resistance applies at AC: input capacitance takes over at high frequency.
Practice
- A sensor has . What input resistance keeps the loading error below ?
- A source drives ; the ammeter has . What does it read, and what is the error?
Answer
- (more exactly ).
- instead of : . The burden voltage is .
Test yourself
Averaging 100 readings reduces which error?
- a calibration offset
- loading
- random noise, by a factor of about 10
- a gross error
A voltmeter of input resistance on a source of reads low by about
- nothing, if the meter is calibrated
Leads left in the A jack while measuring voltage is an example of
- a random error
- a systematic error
- a gross error
- drift
An analog meter rated on its 50 V range has an input resistance of