Home › Electrical & Electronic Measurements › Gross, Systematic, Random and Loading Errors

Gross, Systematic, Random and Loading Errors

A technician measures the midpoint of a 100 kΩ100\ \text{k}\Omega / 100 kΩ100\ \text{k}\Omega divider fed from 10 V10\ \text{V} with an old analog multimeter and reads 4.0 V4.0\ \text{V}. A colleague's DMM reads 4.975 V4.975\ \text{V}. Neither meter is broken, and averaging a thousand readings from either would not bring them together. The difference is a loading error, a systematic error caused by the instrument itself. This lesson sorts measurement errors into four kinds, because each kind has a different cure, and works through loading in detail.

Four kinds of error

The error of a result is the result minus the true value. You never know it exactly (you would not need to measure), but you can know its causes.

Kind Examples Detect it by Cure
gross (blunders) wrong range or function, leads in the A jack, misread scale, digits swapped in the notebook, wrong probe attenuation setting sanity checks, a second method, plotting the data procedure, checklists, automation; never "average out"
systematic calibration offset and gain, loading, burden voltage, thermal EMFs, lead resistance, a temperature coefficient, parallax comparison with a better instrument or a different method; changing the suspected cause calibrate, correct, redesign the method; then assign an uncertainty to the correction
random thermal noise, quantisation, interference that is not synchronous with the reading, operator reading jitter scatter of repeated readings averaging, filtering, longer integration, shielding
drift warm-up, ageing, slow temperature change readings that trend in time warm-up time, recalibration interval, monitoring

The distinction between systematic and random depends on the time scale: 50 Hz50\ \text{Hz} hum is random from reading to reading of a fast scope, but a fixed offset if every reading is synchronised to the mains. The GUM drops the words in favour of "Type A" and "Type B" evaluations (GUM Uncertainty Budgets), but the causes still matter when you decide what to fix.

Averaging removes only random error. The mean of NN readings has a random part that shrinks as 1/N1/\sqrt{N}; the bias does not shrink at all.

python · try it yourself
import numpy as np

rng = np.random.default_rng(7)
true_V = 5.000
bias = 0.012          # meter gain/offset error, V (systematic)
noise = 0.020         # reading-to-reading standard deviation, V (random)

for N in (1, 10, 100, 1000, 10000):
    means = [float(np.mean(true_V + bias + noise * rng.standard_normal(N))) for _ in range(200)]
    spread = float(np.std(means))
    print(f"N = {N:5d}: scatter of the mean {1000 * spread:6.2f} mV, "
          f"mean error {1000 * (np.mean(means) - true_V):+6.2f} mV")
N =     1: scatter of the mean  17.45 mV, mean error  +9.36 mV
N =    10: scatter of the mean   6.86 mV, mean error +11.44 mV
N =   100: scatter of the mean   1.98 mV, mean error +11.95 mV
N =  1000: scatter of the mean   0.65 mV, mean error +12.03 mV
N = 10000: scatter of the mean   0.19 mV, mean error +11.99 mV

The scatter falls by 10\sqrt{10} per decade of NN; the 12 mV12\ \text{mV} bias stays. Past N≈10N \approx 10 the result is limited by the bias, and only calibration helps.

Loading: the instrument becomes part of the circuit

Any circuit seen from the two points you measure between is a Thevenin source VThV_\text{Th} behind RThR_\text{Th} (Thevenin and Norton). A voltmeter of input resistance RinR_\text{in} forms a divider with it:

+ - VTh 5 V RTh 50 kOhm voltmeter Rin 10 MOhm Vm + - + - V 5 V ammeter RA burden R 100 Ohm I burden voltage = I RA
Vm=VTh RinRin+RTh,Vm−VThVTh=−RThRin+RTh≈−RThRinV_m = V_\text{Th}\,\frac{R_\text{in}}{R_\text{in} + R_\text{Th}}, \qquad \frac{V_m - V_\text{Th}}{V_\text{Th}} = -\frac{R_\text{Th}}{R_\text{in} + R_\text{Th}} \approx -\frac{R_\text{Th}}{R_\text{in}}

The relative error is about RTh/RinR_\text{Th}/R_\text{in}: to stay below 0.1 %0.1\,\% the meter needs 10001000 times the source resistance. Because the error is known, it can be corrected:

VTh=Vm(1+RThRin)V_\text{Th} = V_m \left(1 + \frac{R_\text{Th}}{R_\text{in}}\right)

and the uncertainties of RThR_\text{Th} and RinR_\text{in} then enter the budget instead of the whole error.

An ammeter is inserted in series, so its resistance RAR_A (the shunt plus leads and fuse) adds to the loop. A source of VV behind a loop resistance RR gives I=V/RI = V/R without the meter and Im=V/(R+RA)I_m = V/(R + R_A) with it, an error of about −RA/R-R_A/R. The voltage across the meter, IRAI R_A, is its burden voltage (Current Measurement).

At AC the input is an impedance: a 1 MΩ1\ \text{M}\Omega scope input in parallel with 15 pF15\ \text{pF} has ∣Z∣≈1/(2πfC)=10.6 kΩ\lvert Z \rvert \approx 1/(2\pi f C) = 10.6\ \text{k}\Omega at 1 MHz1\ \text{MHz}, so a source that is harmless to load at DC is heavily loaded at radio frequencies (Probes).

Worked example: the divider and two meters

Given: Vs=10.00 VV_s = 10.00\ \text{V}, R1=R2=100 kΩR_1 = R_2 = 100\ \text{k}\Omega; a DMM with Rin=10 MΩR_\text{in} = 10\ \text{M}\Omega; an analog meter rated 20 kΩ/V20\ \text{k}\Omega/\text{V} on its 10 V10\ \text{V} range.

Find: both readings, their errors, and the corrected value from the analog reading.

Solve: VTh=5.000 VV_\text{Th} = 5.000\ \text{V}, RTh=R1∥R2=50 kΩR_\text{Th} = R_1 \parallel R_2 = 50\ \text{k}\Omega. The analog meter's input is 20 kΩ/V×10 V=200 kΩ20\ \text{k}\Omega/\text{V} \times 10\ \text{V} = 200\ \text{k}\Omega.

VDMM=5.000 V×10 MΩ10.05 MΩ=4.975 V(−0.50 %)Vanalog=5.000 V×200 kΩ250 kΩ=4.000 V(−20 %)Vcorrected=4.000 V×(1+50 kΩ200 kΩ)=5.000 V\begin{aligned} V_\text{DMM} &= 5.000\ \text{V} \times \frac{10\ \text{M}\Omega}{10.05\ \text{M}\Omega} = 4.975\ \text{V} && (-0.50\,\%) \\ V_\text{analog} &= 5.000\ \text{V} \times \frac{200\ \text{k}\Omega}{250\ \text{k}\Omega} = 4.000\ \text{V} && (-20\,\%) \\ V_\text{corrected} &= 4.000\ \text{V} \times \left(1 + \frac{50\ \text{k}\Omega}{200\ \text{k}\Omega}\right) = 5.000\ \text{V} \end{aligned}

Check: both readings are low, as loading always lowers a voltage reading on a resistive source. The analog meter's relative error is 0.200/0.00498=400.200/0.00498 = 40 times the DMM's, not 50 (the ratio of the input resistances), because the approximation RTh/RinR_\text{Th}/R_\text{in} fails when the meter is only four times the source resistance. Units cancel in every ratio.

The same deck in ngspice, with the meter as a resistor Rm that the control script changes:

bash · try it yourself
cat > load.cir <<'EOF'
* 100k/100k divider read by voltmeters of different input resistance
V1 in 0 DC 10
R1 in out 100k
R2 out 0 100k
Rm out 0 10meg
.control
op
echo "Rm = 10 Mohm (DMM)"
print v(out)
alter Rm = 1meg
op
echo "Rm = 1 Mohm (scope input)"
print v(out)
alter Rm = 200k
op
echo "Rm = 200 kohm (analog meter, 20 kohm/V on 10 V)"
print v(out)
.endc
.end
EOF
ngspice -b load.cir 2>&1 | grep -E '^(v\(|Rm)'
Rm = 10 Mohm (DMM)
v(out) = 4.975124e+00
Rm = 1 Mohm (scope input)
v(out) = 4.761905e+00
Rm = 200 kohm (analog meter, 20 kohm/V on 10 V)
v(out) = 4.000000e+00

Guided lab: how high must the meter's resistance be?

Hand calculation. For the divider above, the error is −50k/(Rin+50k)-50\text{k}/(R_\text{in} + 50\text{k}): −20 %-20\,\% at 200 kΩ200\ \text{k}\Omega, −4.8 %-4.8\,\% at 1 MΩ1\ \text{M}\Omega, −0.50 %-0.50\,\% at 10 MΩ10\ \text{M}\Omega.

Simulation. Rm is the meter. Slide it from 100 kΩ100\ \text{k}\Omega to 10 MΩ10\ \text{M}\Omega and read node out: 3.333 V3.333\ \text{V} at 100 kΩ100\ \text{k}\Omega, 4.762 V4.762\ \text{V} at 1 MΩ1\ \text{M}\Omega, 4.975 V4.975\ \text{V} at 10 MΩ10\ \text{M}\Omega. A 1 kΩ1\ \text{k}\Omega / 1 kΩ1\ \text{k}\Omega divider (RTh=500 ΩR_\text{Th} = 500\ \Omega) would err by only 0.5 %0.5\,\% at 100 kΩ100\ \text{k}\Omega and 0.005 %0.005\,\% at 10 MΩ10\ \text{M}\Omega: low source resistance is the real cure.

node voltages (to ground 0)in10.00 Vout4.975 VnetlistV1 in 0 DC 10R1 in out 100kR2 out 0 100kRm out 0 10megelement currents (node 1 to node 2 through it)V1in>0-50.2 uAR1in>out50.2 uAR2out>049.8 uARmout>0498 nA

What you would measure. With 1 %1\,\% resistors and a 10 V10\ \text{V} supply set with the DMM itself: a DMM reading of about 4.975 V4.975\ \text{V} (within ±0.05 V\pm 0.05\ \text{V} from resistor tolerance), a scope reading near 4.76 V4.76\ \text{V} (if your scope has 1 MΩ1\ \text{M}\Omega input and a 1x probe; a 10x probe gives 10 MΩ10\ \text{M}\Omega and about 4.97 V4.97\ \text{V}). Some DMMs switch to a >10 GΩ>10\ \text{G}\Omega input on their lowest DC ranges; on those the loading disappears, which is itself a useful check of the input-resistance setting.

Hazards and safe setup. Current-limited 10 V10\ \text{V} supply; the divider draws 50 μA50\ \mu\text{A}. No hazard beyond the usual: the scope's probe ground goes to circuit ground only.

Pitfalls

  • Averaging to "improve accuracy" when the error is systematic.
  • Forgetting that the input resistance changes with range and function (analog meters: per volt of range; some DMMs: 10 MΩ10\ \text{M}\Omega or >10 GΩ>10\ \text{G}\Omega selectable).
  • Correcting for loading but leaving the correction out of the uncertainty budget.
  • Assuming a DC input resistance applies at AC: input capacitance takes over at high frequency.

Practice

  1. A sensor has RTh=2 kΩR_\text{Th} = 2\ \text{k}\Omega. What input resistance keeps the loading error below 0.01 %0.01\,\%?
  2. A 5 V5\ \text{V} source drives 100 Ω100\ \Omega; the ammeter has RA=2.5 ΩR_A = 2.5\ \Omega. What does it read, and what is the error?
Answer
  1. Rin≥RTh/10−4=20 MΩR_\text{in} \ge R_\text{Th}/10^{-4} = 20\ \text{M}\Omega (more exactly 19.998 MΩ19.998\ \text{M}\Omega).
  2. 5/102.5=48.78 mA5/102.5 = 48.78\ \text{mA} instead of 50.00 mA50.00\ \text{mA}: −2.4 %-2.4\,\%. The burden voltage is 48.78 mA×2.5 Ω=0.12 V48.78\ \text{mA} \times 2.5\ \Omega = 0.12\ \text{V}.

Test yourself

Averaging 100 readings reduces which error?

  1. a calibration offset
  2. loading
  3. random noise, by a factor of about 10
  4. a gross error

A voltmeter of input resistance RinR_\text{in} on a source of RThR_\text{Th} reads low by about

  1. Rin/RThR_\text{in}/R_\text{Th}
  2. RTh/RinR_\text{Th}/R_\text{in}
  3. (RTh/Rin)2(R_\text{Th}/R_\text{in})^2
  4. nothing, if the meter is calibrated

Leads left in the A jack while measuring voltage is an example of

  1. a random error
  2. a systematic error
  3. a gross error
  4. drift

An analog meter rated 20 kΩ/V20\ \text{k}\Omega/\text{V} on its 50 V range has an input resistance of

  1. 20 kΩ20\ \text{k}\Omega
  2. 1 MΩ1\ \text{M}\Omega
  3. 2.5 kΩ2.5\ \text{k}\Omega
  4. 10 MΩ10\ \text{M}\Omega