Home › Electrical & Electronic Measurements › Analog Meters: PMMC, Moving Iron, Shunts and Multipliers

Analog Meters: PMMC, Moving Iron, Shunts and Multipliers

Before the DMM, every voltage and current on a bench was read from a needle. Analog meters survive in panel instruments, in cheap multimeters, as the trend display on many DMMs, and in the vocabulary: "ohms per volt", "shunt", "multiplier", "full-scale deflection". They are also the clearest example of a principle every instrument shares: a sensing element with a small, fixed range, extended by passive networks whose accuracy and temperature behaviour set the accuracy of the whole meter. This lesson treats them briefly and historically, then designs the range networks, which reappear unchanged in DMM front ends and current sensors.

The permanent-magnet moving-coil (PMMC) movement

A light coil of NN turns and area AA hangs in the radial field BB of a permanent magnet. A current II produces a torque NBAIN B A I; a hairspring opposes with kθk\theta. At rest

θ=NBAk I\theta = \frac{N B A}{k}\, I

so the deflection is linear in current and the scale is uniform. The movement responds to the average of the current (the inertia of the coil filters anything faster than a few hertz), so it reads zero on a symmetric AC signal; AC ranges add a rectifier and a scale calibrated in RMS for a sine, the subject of True RMS. Typical movements need 50 μA50\ \mu\text{A} to 1 mA1\ \text{mA} for full-scale deflection (FSD), with coil resistances of hundreds to a few thousand ohms. Accuracy classes are given as a percentage of the fiducial value, usually the full-scale value (IEC 60051-1:2016: class indices from 0.05 to 5, such as 0.5, 1, 1.5, 2.5), so a reading at one third of the scale has three times the class figure in relative error.

Moving-iron meters attract or repel a soft-iron vane with the field of a fixed coil. The force goes as I2I^2, so they respond to the mean square and read the RMS value of any waveform within their (low) bandwidth, with a cramped, non-linear scale. Robust and cheap, they were the standard mains panel ammeter and voltmeter. Electrodynamometer movements replace the magnet by a second coil; with one coil carrying the load current and the other a current proportional to the voltage, the torque follows the instantaneous product v iv\,i, which makes them wattmeters (Power Measurement).

Extending the range

G movement Rm 2 kOhm Rsh 10.05 Ohm I Im I - Im + - ammeter: shunt in parallel + Rs 198 kOhm G 50 uA - Rm = 2 kOhm voltmeter: multiplier in series Rin = 200 kOhm

Ammeter: a shunt in parallel. To read II at full scale with a movement of FSD current ImI_m and resistance RmR_m, the shunt must carry I−ImI - I_m at the same voltage ImRmI_m R_m:

Rsh=ImRmI−ImR_\text{sh} = \frac{I_m R_m}{I - I_m}

Voltmeter: a multiplier in series. For a full-scale voltage VV:

Rs=VIm−Rm,Rin=Rs+Rm=VImR_s = \frac{V}{I_m} - R_m, \qquad R_\text{in} = R_s + R_m = \frac{V}{I_m}

The input resistance is proportional to the range, so a voltmeter is characterised by its sensitivity 1/Im1/I_m in ohms per volt: 20 kΩ/V20\ \text{k}\Omega/\text{V} for a 50 μA50\ \mu\text{A} movement. That is the origin of the loading error in the previous section.

Worked example: ranges for a 50 microamp movement

Given: a PMMC movement with Im=50 μAI_m = 50\ \mu\text{A} and Rm=2000 ΩR_m = 2000\ \Omega.

Find: shunts for 1 mA1\ \text{mA} to 1 A1\ \text{A}, multipliers for 11 to 100 V100\ \text{V}, and the effect of a 10 K10\ \text{K} rise of the copper coil's temperature on the 1 A1\ \text{A} range.

Solve: the full-scale voltage is ImRm=100 mVI_m R_m = 100\ \text{mV}. For 10 mA10\ \text{mA}: Rsh=0.1 V/9.95 mA=10.05 ΩR_\text{sh} = 0.1\ \text{V}/9.95\ \text{mA} = 10.05\ \Omega. For 10 V10\ \text{V}: Rs=10 V/50 μA−2 kΩ=198 kΩR_s = 10\ \text{V}/50\ \mu\text{A} - 2\ \text{k}\Omega = 198\ \text{k}\Omega. Copper's resistance rises by about 0.393 %/K0.393\,\%/\text{K}, so 10 K10\ \text{K} adds 3.9 %3.9\,\% to RmR_m while the manganin shunt stays constant; the coil's share of the current falls by almost the same fraction.

python · try it yourself
Im, Rm = 50e-6, 2000.0             # full-scale current and coil resistance of the movement
Vm = Im * Rm                        # voltage across the movement at full scale
print(f"movement: {1e6 * Im:.0f} uA full scale, {Rm:.0f} ohm, {1000 * Vm:.0f} mV, {1 / Im / 1000:.0f} kohm/V")

print("ammeter ranges (shunt in parallel):")
for I in (1e-3, 10e-3, 100e-3, 1.0):
    Rsh = Vm / (I - Im)
    print(f"  {I:7.3f} A: Rsh = {Rsh:9.4f} ohm, meter resistance {Rsh * Rm / (Rsh + Rm):8.4f} ohm")

print("voltmeter ranges (multiplier in series):")
for V in (1.0, 10.0, 100.0):
    Rs = V / Im - Rm
    print(f"  {V:5.0f} V: Rs = {Rs / 1000:8.1f} kohm, input resistance {(Rs + Rm) / 1000:7.0f} kohm")

# Temperature: copper coil +0.393 %/K, manganin shunt ~0; 10 K warmer
Rm_hot = Rm * (1 + 0.00393 * 10)
Rsh = Vm / (1.0 - Im)
err = (Rsh / (Rsh + Rm_hot)) / (Rsh / (Rsh + Rm)) - 1
print(f"1 A range, coil 10 K warmer: reading error {100 * err:+.2f} %")
Rswamp = 3 * Rm                     # manganin swamping resistor, 3x the coil, in series with the coil
Rsh2 = Im * (Rm + Rswamp) / (1.0 - Im)
err2 = ((Rsh2 / (Rsh2 + Rm_hot + Rswamp)) / (Rsh2 / (Rsh2 + Rm + Rswamp))) - 1
print(f"with a 6 kohm manganin swamping resistor: {100 * err2:+.2f} %")
movement: 50 uA full scale, 2000 ohm, 100 mV, 20 kohm/V
ammeter ranges (shunt in parallel):
    0.001 A: Rsh =  105.2632 ohm, meter resistance 100.0000 ohm
    0.010 A: Rsh =   10.0503 ohm, meter resistance  10.0000 ohm
    0.100 A: Rsh =    1.0005 ohm, meter resistance   1.0000 ohm
    1.000 A: Rsh =    0.1000 ohm, meter resistance   0.1000 ohm
voltmeter ranges (multiplier in series):
      1 V: Rs =     18.0 kohm, input resistance      20 kohm
     10 V: Rs =    198.0 kohm, input resistance     200 kohm
    100 V: Rs =   1998.0 kohm, input resistance    2000 kohm
1 A range, coil 10 K warmer: reading error -3.78 %
with a 6 kohm manganin swamping resistor: -0.97 %

Check: every ammeter range presents 100 mV100\ \text{mV} at full scale (the meter resistance equals 0.1 V/I0.1\ \text{V}/I), and every voltmeter range 20 kΩ/V20\ \text{k}\Omega/\text{V}. The fix for the temperature error, a swamping resistor of manganin in series with the coil, cuts the error from −3.8 %-3.8\,\% to −1.0 %-1.0\,\% at the cost of a four times larger full-scale drop (400 mV400\ \text{mV}), a trade between accuracy and burden voltage that every current meter makes.

Guided lab: the shunt divides the current

Hand calculation. On the 10 mA10\ \text{mA} range (Rsh=10.05 ΩR_\text{sh} = 10.05\ \Omega), a 10 mA10\ \text{mA} input sends 10 mA×10.05/2010.05=50.0 μA10\ \text{mA} \times 10.05/2010.05 = 50.0\ \mu\text{A} through the movement, and 5 mA5\ \text{mA} sends 25.0 μA25.0\ \mu\text{A} (half scale): the division ratio is constant, so the scale stays linear.

Simulation. Rm is the coil, Rsh the shunt and I1 the current being measured. At the default the lab shows I(Rm) =50.0 μA= 50.0\ \mu\text{A} and node a at 100 mV100\ \text{mV}. Slide I1 and watch the movement current follow in proportion; then slide Rsh to see how a 1 %1\,\% shunt error becomes a 1 %1\,\% reading error.

node voltages (to ground 0)a100.0 mVnetlistI1 0 a DC 10mRm a 0 2kRsh a 0 10.05element currents (node 1 to node 2 through it)I10>a10.0 mARma>050.0 uARsha>09.95 mA

What you would measure. If you have an analog panel meter (often 100 μA100\ \mu\text{A} or 1 mA1\ \text{mA} FSD), measure its coil resistance with a DMM on a high ohms range (the DMM's test current may exceed FSD on low ranges; check its open-circuit current first), then build a shunt from a resistor in parallel and compare the needle with a DMM in series. Expect agreement within the meter's class (for class 2.5, ±2.5 %\pm 2.5\,\% of full scale) and a visible error if the shunt's tolerance is 5 %5\,\%.

Hazards and safe setup. Feed the meter from a current-limited supply through a series resistor that limits the current to the range (for example 5 V5\ \text{V} through 560 Ω560\ \Omega for about 9 mA9\ \text{mA}). Never connect a bare movement across a supply: 100 mV100\ \text{mV} already drives full scale.

Pitfalls

  • Reading a class 1.5 meter at 10 %10\,\% of scale and expecting 1.5 %1.5\,\% accuracy (15 %15\,\% is the honest figure).
  • Forgetting that a rectifier AC meter is calibrated for sine waves only.
  • Making shunts from copper: the temperature coefficient spoils the range. Manganin and similar alloys exist for this reason.
  • Parallax: read with your eye above the mirror strip so the needle hides its reflection.

Practice

  1. A 1 mA1\ \text{mA}, 100 Ω100\ \Omega movement must read 5 A5\ \text{A}. What shunt, and what is the meter's burden voltage at full scale?
  2. What is the input resistance of a 1 mA1\ \text{mA} movement used as a 50 V50\ \text{V} voltmeter?
Answer
  1. Rsh=0.1 V/4.999 A=20.004 mΩR_\text{sh} = 0.1\ \text{V}/4.999\ \text{A} = 20.004\ \text{m}\Omega; burden 100 mV100\ \text{mV} (plus lead and fuse drops).
  2. 50 V/1 mA=50 kΩ50\ \text{V}/1\ \text{mA} = 50\ \text{k}\Omega (1 kΩ/V1\ \text{k}\Omega/\text{V}): a heavy load on most electronic circuits.

Test yourself

A PMMC movement responds to

  1. the RMS value
  2. the peak value
  3. the average value
  4. the square of the current

A voltmeter rated 20 kΩ/V20\ \text{k}\Omega/\text{V} uses a movement with a full-scale current of

  1. 20 μA20\ \mu\text{A}
  2. 50 μA50\ \mu\text{A}
  3. 1 mA1\ \text{mA}
  4. 20 mA20\ \text{mA}

Why is a moving-iron meter's scale non-linear?

  1. the spring is non-linear
  2. its torque is proportional to I2I^2
  3. it uses a rectifier
  4. the magnet saturates

A swamping resistor in series with the coil

  1. increases the sensitivity
  2. reduces the effect of the coil's temperature coefficient
  3. removes the need for a shunt
  4. lowers the burden voltage