Analog Meters: PMMC, Moving Iron, Shunts and Multipliers
Before the DMM, every voltage and current on a bench was read from a needle. Analog meters survive in panel instruments, in cheap multimeters, as the trend display on many DMMs, and in the vocabulary: "ohms per volt", "shunt", "multiplier", "full-scale deflection". They are also the clearest example of a principle every instrument shares: a sensing element with a small, fixed range, extended by passive networks whose accuracy and temperature behaviour set the accuracy of the whole meter. This lesson treats them briefly and historically, then designs the range networks, which reappear unchanged in DMM front ends and current sensors.
The permanent-magnet moving-coil (PMMC) movement
A light coil of turns and area hangs in the radial field of a permanent magnet. A current produces a torque ; a hairspring opposes with . At rest
so the deflection is linear in current and the scale is uniform. The movement responds to the average of the current (the inertia of the coil filters anything faster than a few hertz), so it reads zero on a symmetric AC signal; AC ranges add a rectifier and a scale calibrated in RMS for a sine, the subject of True RMS. Typical movements need to for full-scale deflection (FSD), with coil resistances of hundreds to a few thousand ohms. Accuracy classes are given as a percentage of the fiducial value, usually the full-scale value (IEC 60051-1:2016: class indices from 0.05 to 5, such as 0.5, 1, 1.5, 2.5), so a reading at one third of the scale has three times the class figure in relative error.
Moving-iron meters attract or repel a soft-iron vane with the field of a fixed coil. The force goes as , so they respond to the mean square and read the RMS value of any waveform within their (low) bandwidth, with a cramped, non-linear scale. Robust and cheap, they were the standard mains panel ammeter and voltmeter. Electrodynamometer movements replace the magnet by a second coil; with one coil carrying the load current and the other a current proportional to the voltage, the torque follows the instantaneous product , which makes them wattmeters (Power Measurement).
Extending the range
Ammeter: a shunt in parallel. To read at full scale with a movement of FSD current and resistance , the shunt must carry at the same voltage :
Voltmeter: a multiplier in series. For a full-scale voltage :
The input resistance is proportional to the range, so a voltmeter is characterised by its sensitivity in ohms per volt: for a movement. That is the origin of the loading error in the previous section.
Worked example: ranges for a 50 microamp movement
Given: a PMMC movement with and .
Find: shunts for to , multipliers for to , and the effect of a rise of the copper coil's temperature on the range.
Solve: the full-scale voltage is . For : . For : . Copper's resistance rises by about , so adds to while the manganin shunt stays constant; the coil's share of the current falls by almost the same fraction.
Im, Rm = 50e-6, 2000.0 # full-scale current and coil resistance of the movement
Vm = Im * Rm # voltage across the movement at full scale
print(f"movement: {1e6 * Im:.0f} uA full scale, {Rm:.0f} ohm, {1000 * Vm:.0f} mV, {1 / Im / 1000:.0f} kohm/V")
print("ammeter ranges (shunt in parallel):")
for I in (1e-3, 10e-3, 100e-3, 1.0):
Rsh = Vm / (I - Im)
print(f" {I:7.3f} A: Rsh = {Rsh:9.4f} ohm, meter resistance {Rsh * Rm / (Rsh + Rm):8.4f} ohm")
print("voltmeter ranges (multiplier in series):")
for V in (1.0, 10.0, 100.0):
Rs = V / Im - Rm
print(f" {V:5.0f} V: Rs = {Rs / 1000:8.1f} kohm, input resistance {(Rs + Rm) / 1000:7.0f} kohm")
# Temperature: copper coil +0.393 %/K, manganin shunt ~0; 10 K warmer
Rm_hot = Rm * (1 + 0.00393 * 10)
Rsh = Vm / (1.0 - Im)
err = (Rsh / (Rsh + Rm_hot)) / (Rsh / (Rsh + Rm)) - 1
print(f"1 A range, coil 10 K warmer: reading error {100 * err:+.2f} %")
Rswamp = 3 * Rm # manganin swamping resistor, 3x the coil, in series with the coil
Rsh2 = Im * (Rm + Rswamp) / (1.0 - Im)
err2 = ((Rsh2 / (Rsh2 + Rm_hot + Rswamp)) / (Rsh2 / (Rsh2 + Rm + Rswamp))) - 1
print(f"with a 6 kohm manganin swamping resistor: {100 * err2:+.2f} %")movement: 50 uA full scale, 2000 ohm, 100 mV, 20 kohm/V
ammeter ranges (shunt in parallel):
0.001 A: Rsh = 105.2632 ohm, meter resistance 100.0000 ohm
0.010 A: Rsh = 10.0503 ohm, meter resistance 10.0000 ohm
0.100 A: Rsh = 1.0005 ohm, meter resistance 1.0000 ohm
1.000 A: Rsh = 0.1000 ohm, meter resistance 0.1000 ohm
voltmeter ranges (multiplier in series):
1 V: Rs = 18.0 kohm, input resistance 20 kohm
10 V: Rs = 198.0 kohm, input resistance 200 kohm
100 V: Rs = 1998.0 kohm, input resistance 2000 kohm
1 A range, coil 10 K warmer: reading error -3.78 %
with a 6 kohm manganin swamping resistor: -0.97 %
Check: every ammeter range presents at full scale (the meter resistance equals ), and every voltmeter range . The fix for the temperature error, a swamping resistor of manganin in series with the coil, cuts the error from to at the cost of a four times larger full-scale drop (), a trade between accuracy and burden voltage that every current meter makes.
Guided lab: the shunt divides the current
Hand calculation. On the range (), a input sends through the movement, and sends (half scale): the division ratio is constant, so the scale stays linear.
Simulation. Rm is the coil, Rsh the shunt and I1 the current being measured. At the default the lab shows I(Rm) and node a at . Slide I1 and watch the movement current follow in proportion; then slide Rsh to see how a shunt error becomes a reading error.
What you would measure. If you have an analog panel meter (often or FSD), measure its coil resistance with a DMM on a high ohms range (the DMM's test current may exceed FSD on low ranges; check its open-circuit current first), then build a shunt from a resistor in parallel and compare the needle with a DMM in series. Expect agreement within the meter's class (for class 2.5, of full scale) and a visible error if the shunt's tolerance is .
Hazards and safe setup. Feed the meter from a current-limited supply through a series resistor that limits the current to the range (for example through for about ). Never connect a bare movement across a supply: already drives full scale.
Pitfalls
- Reading a class 1.5 meter at of scale and expecting accuracy ( is the honest figure).
- Forgetting that a rectifier AC meter is calibrated for sine waves only.
- Making shunts from copper: the temperature coefficient spoils the range. Manganin and similar alloys exist for this reason.
- Parallax: read with your eye above the mirror strip so the needle hides its reflection.
Practice
- A , movement must read . What shunt, and what is the meter's burden voltage at full scale?
- What is the input resistance of a movement used as a voltmeter?
Answer
- ; burden (plus lead and fuse drops).
- (): a heavy load on most electronic circuits.
Test yourself
A PMMC movement responds to
- the RMS value
- the peak value
- the average value
- the square of the current
A voltmeter rated uses a movement with a full-scale current of
Why is a moving-iron meter's scale non-linear?
- the spring is non-linear
- its torque is proportional to
- it uses a rectifier
- the magnet saturates
A swamping resistor in series with the coil
- increases the sensitivity
- reduces the effect of the coil's temperature coefficient
- removes the need for a shunt
- lowers the burden voltage