LCR Meters and Impedance Analysers
A bench LCR meter answers "what is this component?" in one press: , at . But change the test frequency to and the same capacitor reads in series mode and in parallel mode; at it reads nonsense. Nothing is wrong with the meter. A real component is a network, and the meter reports the parameters of whichever two-element model you select at the frequency you chose. This lesson explains how the meter measures impedance, how to pick the model, frequency and connection, and how compensation removes the fixture.
Measuring impedance: the auto-balancing bridge
Modern LCR meters (to a few MHz) use the auto-balancing bridge: a sinusoidal source drives the device under test (DUT) from the High side; the Low side goes to the virtual-ground input of a current-to-voltage amplifier, which forces the Low terminal to and outputs . Two vector (phase-sensitive) voltmeters measure the voltage across the DUT and , and the processor computes
as a complex number. It is still a bridge in spirit (the amplifier "balances" the Low terminal to zero), and still a ratio measurement: the source level cancels.
The terminals come as four-terminal pairs: Hcur and Lcur carry the current, Hpot and Lpot sense the voltage, each with its coaxial shield. This is the Kelvin connection extended to AC: it removes lead impedance from milliohm and microhenry measurements. Above a few MHz, RF I-V and network analyser methods take over (Spectrum and Network Analysers); an impedance analyser is an LCR meter that sweeps frequency, DC bias and level and plots the result.
From |Z| and theta to C, L, R, D and Q
The meter knows only at one frequency. It reports it through a two-element model:
| Model | Parameters | Relations () |
|---|---|---|
| series C + R | , (ESR) | , |
| parallel | , | |
| series L + R | , | |
| parallel | , |
with the dissipation factor and quality factor
Both models describe the same , so they agree when and diverge when the loss is large. Choose the model that matches the dominant physics: low impedances (large capacitors, small inductors, below about ) in series mode, because the series resistance (ESR, winding resistance) matters most; high impedances (small capacitors, large inductors, above about ) in parallel mode, because leakage or core loss in parallel matters most.
Test conditions matter. Electrolytic capacitors are specified at (rectifier ripple), film and ceramic capacitors at (Class II ceramics above : ; Class I ceramics up to : ), inductors at their working frequency. Class II ceramic capacitors (X7R, X5R) lose capacitance with DC bias (often more than half at rated voltage) and change with AC test level, so a datasheet value is only reproducible at the stated level, typically or . Every component has a self-resonant frequency: above it a capacitor behaves as an inductor.
Worked example: a 10 microfarad ceramic capacitor across frequency
Given: with and in series.
Find: the self-resonant frequency and the series and parallel readings at .
Solve:
Check: is slightly above because the ESL cancels part of the capacitive reactance; as approaches the apparent grows without limit, and above it the reactance is inductive. The program sweeps the same model and also shows open/short compensation:
import numpy as np
def lcr_readings(Z, f):
"""What an LCR meter displays for impedance Z at frequency f (series and parallel models)."""
w = 2 * np.pi * f
Y = 1 / Z
if Z.imag < 0: # capacitive
Cs, Rs = -1 / (w * Z.imag), Z.real
Cp, Rp = Y.imag / w, 1 / Y.real
return f"Cs = {Cs * 1e6:8.3f} uF ESR = {1e3 * Rs:6.1f} mohm Cp = {Cp * 1e6:8.3f} uF D = {Rs / -Z.imag:.4f}"
Ls, Rs = Z.imag / w, Z.real
return f"Ls = {Ls * 1e9:8.2f} nH Rs = {1e3 * Rs:6.1f} mohm (inductive above self-resonance)"
# A 10 uF ceramic capacitor modelled as C + ESR + ESL in series
C, ESR, ESL = 10e-6, 0.050, 2e-9
f_sr = 1 / (2 * np.pi * np.sqrt(ESL * C))
print(f"self-resonance {f_sr / 1e6:.3f} MHz")
for f in (120.0, 1e3, 100e3, 1e6, 10e6):
w = 2 * np.pi * f
Z = ESR + 1j * w * ESL + 1 / (1j * w * C)
print(f"{f:>10.0f} Hz: |Z| = {abs(Z):9.4f} ohm, theta = {np.degrees(np.angle(Z)):+6.1f} deg | {lcr_readings(Z, f)}")
# Open/short compensation: fixture with 5 mohm + 10 nH in series and 2 pF stray in parallel at 1 MHz
f = 1e6
w = 2 * np.pi * f
Zs = 0.005 + 1j * w * 10e-9 # measured with the fixture shorted
Yo = 1j * w * 2e-12 # measured with the fixture open
Zdut = 1 / (1j * w * 100e-12) # true DUT: 100 pF
Zm = 1 / (1 / (Zdut + Zs) + Yo) # what the meter sees
Zc = (Zm - Zs) / (1 - (Zm - Zs) * Yo) # compensated
for name, Zx in (("uncompensated", Zm), ("compensated", Zc)):
print(f"{name:14s}: Cs = {-1 / (w * Zx.imag) * 1e12:7.3f} pF")self-resonance 1.125 MHz
120 Hz: |Z| = 132.6291 ohm, theta = -90.0 deg | Cs = 10.000 uF ESR = 50.0 mohm Cp = 10.000 uF D = 0.0004
1000 Hz: |Z| = 15.9156 ohm, theta = -89.8 deg | Cs = 10.000 uF ESR = 50.0 mohm Cp = 10.000 uF D = 0.0031
100000 Hz: |Z| = 0.1656 ohm, theta = -72.4 deg | Cs = 10.080 uF ESR = 50.0 mohm Cp = 9.161 uF D = 0.3167
1000000 Hz: |Z| = 0.0501 ohm, theta = -3.8 deg | Cs = 47.521 uF ESR = 50.0 mohm Cp = 0.212 uF D = 14.9293
10000000 Hz: |Z| = 0.1338 ohm, theta = +68.1 deg | Ls = 1.97 nH Rs = 50.0 mohm (inductive above self-resonance)
uncompensated : Cs = 102.004 pF
compensated : Cs = 100.000 pF
Compensation: removing the fixture
Test leads and fixtures add a series impedance (lead resistance and inductance) and a parallel admittance (stray capacitance between the terminals). The meter measures them with the fixture shorted and open, and corrects every later reading:
In the program a stray turned into ; compensation restores . Perform open/short compensation at every test frequency you use, with the fixture exactly as it will be (cable positions included); a load compensation against a known standard removes the remaining errors.
Guided lab: impedance of a real capacitor
Hand calculation. As in the worked example: at , a minimum of (the ESR) at , inductive above.
Simulation. A AC current source drives the capacitor model, so the voltage at node a is the impedance. At the readouts show at , the and of the program. Move the frequency slider to about : the phase passes through and bottoms out at the ESR; above it the phase is positive (inductive). Slide the ESL (lead length: roughly per millimetre of lead or trace) and watch the resonance move.
What you would measure. On an LCR meter (or a DMM capacitance range for the value only), measure a X7R MLCC rated at , : expect within (the usual tolerance) and around to . If your meter has a DC bias function, apply : expect the capacitance to drop by roughly to (it depends strongly on the part; check its datasheet curve). Repeat at in series and parallel mode and compare with the model.
Hazards and safe setup. LCR meter test levels are around . If you use external DC bias above a few volts, discharge the capacitor before removing it. Never put a charged capacitor on the meter's terminals.
Pitfalls
- Reporting a capacitance without its test frequency, level and bias.
- Choosing the series model for a high-impedance part (or the reverse) and quoting a value that depends on the loss.
- Skipping open/short compensation, especially for small capacitances and inductances.
- Measuring above a component's self-resonance and believing the "C" reading.
Practice
- An LCR meter shows , at . Find , and .
- A capacitor has and . What does the parallel model report?
Answer
- , , , .
- .
Test yourself
In an auto-balancing bridge the Low terminal is held at
- the source voltage
- virtual ground by the current-to-voltage amplifier
- half the supply
- earth through a resistor
For a capacitor at () the better model is
- series
- parallel
- either, always identical
- inductive
Above its self-resonant frequency a capacitor
- has zero impedance
- behaves as an inductor
- has infinite D at all frequencies
- reads its nominal value
Open/short compensation corrects for
- the DUT's tolerance
- lead and fixture series impedance and stray parallel admittance
- DC bias effects
- temperature